Tumor blood flow measured by perfusion computed tomography and 15O-labeled water positron emission tomography: a comparison study.

Journal Article (Clinical Trial;Journal Article)

OBJECTIVES: To compare blood flow measurements of tumors assessed by perfusion computed tomography (pCT) and the clinical gold standard of 15O-labeled water positron emission tomography (15O-PET). METHODS: Blood flows were estimated by pCT (4-row multidetector, CT Perfusion 3.1) and 15O-PET (Posicam, first-pass model) in 14 patients with solid tumors, totaling 22 index tumors and 57 matched pairs of examinations. Blood flow estimates were compared using t test, Bland-Altman, and linear mixed regression analyses. RESULTS: There was no significant difference between the mean (SD) blood flow values measured by pCT and 15O-PET: 25.9 +/- 15.4 and 27.8 +/- 14.0 mL/min per 100 g, respectively. CONCLUSIONS: The demonstration of a good correlation between pCT and 15O-PET potentially enables the use of pCT, which is more widely available than 15O-PET, when tumor blood flow estimates are required, particularly in the context of clinical studies.

Full Text

Duke Authors

Cited Authors

  • Ng, CS; Kodama, Y; Mullani, NA; Barron, BJ; Wei, W; Herbst, RS; Abbruzzese, JL; Charnsangavej, C

Published Date

  • 2009

Published In

Volume / Issue

  • 33 / 3

Start / End Page

  • 460 - 465

PubMed ID

  • 19478644

Electronic International Standard Serial Number (EISSN)

  • 1532-3145

Digital Object Identifier (DOI)

  • 10.1097/RCT.0b013e318182d2e0

Language

  • eng

Conference Location