Skip to main content
Journal cover image

What observer models best reflect low-contrast detectability in CT?

Publication ,  Conference
Solomon, J; Samei, E
Published in: Progress in Biomedical Optics and Imaging - Proceedings of SPIE
January 1, 2015

The purpose of this work was to compare CT low-contrast detectability as measured via human perception experiments with observer model surrogates of image quality measured directly from the images. A phantom was designed with a range of low-contrast circular inserts representing 5 contrast levels and 3 sizes. The phantom was imaged repeatedly (20 times) on a third-generation dual-source CT scanner (SOMATOM Definition Force, Siemens Healthcare). Images were reconstructed at 0.6 mm slice thickness using filtered back projection (FBP) and advanced modeled iterative reconstruction (ADMIRE) and were assessed by eleven blinded and independent readers using a two alternative forced choice (2AFC) detection experiment. The human scores were taken as the accuracy, averaged across observers. The predicted performance was computed directly from the images for several traditional image quality metrics and model observers including contrast to noise ratio (CNR), area weighted CNR (CNRa), non-prewhitening matched filter (NPW), non-prewhitening matched filter with an eye filter (NPWE), channelized Hotelling observer (CHO), and channelized Hotelling observer with internal noise (CHOi). The correlation between model observer predictions and human performance was assessed using linear regression analysis. The coefficient of determination (R2) was used as goodness-of-fit metric to determine how well each model observer predicts human performance. R2 was 0.11, 0.71, 0.73, 0.77, 0.60, and 0.72 for CNR, CNRa, NPW, NPWE, CHO, and CHOi, respectively. The findings demonstrate NPW, NPWE, and CHOi all to have strong correlation with human performance and could be used to optimize scan and reconstruction settings.

Duke Scholars

Published In

Progress in Biomedical Optics and Imaging - Proceedings of SPIE

DOI

ISSN

1605-7422

ISBN

9781628415063

Publication Date

January 1, 2015

Volume

9416
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Solomon, J., & Samei, E. (2015). What observer models best reflect low-contrast detectability in CT? In Progress in Biomedical Optics and Imaging - Proceedings of SPIE (Vol. 9416). https://doi.org/10.1117/12.2081655
Solomon, J., and E. Samei. “What observer models best reflect low-contrast detectability in CT?” In Progress in Biomedical Optics and Imaging - Proceedings of SPIE, Vol. 9416, 2015. https://doi.org/10.1117/12.2081655.
Solomon J, Samei E. What observer models best reflect low-contrast detectability in CT? In: Progress in Biomedical Optics and Imaging - Proceedings of SPIE. 2015.
Solomon, J., and E. Samei. “What observer models best reflect low-contrast detectability in CT?Progress in Biomedical Optics and Imaging - Proceedings of SPIE, vol. 9416, 2015. Scopus, doi:10.1117/12.2081655.
Solomon J, Samei E. What observer models best reflect low-contrast detectability in CT? Progress in Biomedical Optics and Imaging - Proceedings of SPIE. 2015.
Journal cover image

Published In

Progress in Biomedical Optics and Imaging - Proceedings of SPIE

DOI

ISSN

1605-7422

ISBN

9781628415063

Publication Date

January 1, 2015

Volume

9416