Skip to main content
Journal cover image

Cost-effective, Reusable, Leak-resistant Ultrasound-guided Vascular Access Trainer.

Publication ,  Journal Article
Morrow, DS; Broder, J
Published in: J Emerg Med
September 2015

BACKGROUND: Ultrasound guidance for insertion of a peripheral venous catheter is becoming common practice in many emergency departments in the difficult-to-access patient, and simulation has become an important tool for health care practitioners to learn this technique. Commercial trainers are expensive, and low-cost alternatives described to date provide a sub-optimal training experience. We introduce ballistics gel as a new material for the creation of simulating phantoms. MATERIALS AND METHODS: Directions describe construction of a simulating phantom composed of 10% ballistic gelatin and commonly available latex tubing. The model's success as used by one residency training program and medical school is described. RESULTS: Cost per phantom was $22.83, with less than an hour preparation time per phantom. We found these phantoms to offer a comparable user experience to commercially available products and better than other homemade products. DISCUSSION: Ballistics gel is a novel material for production of simulation phantoms that provides a low-cost, realistic simulation experience. The clear gel material works well for novice learners, and opacifying agents can be added to increase difficulty for more advanced learners. The material offers flexibility in design to make models for a broad range of skill instruction. CONCLUSION: A relatively quick and easy process using ballistics gel allowed us to create a simulation experience similar to commercially available trainers at a fraction of the cost.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

J Emerg Med

DOI

ISSN

0736-4679

Publication Date

September 2015

Volume

49

Issue

3

Start / End Page

313 / 317

Location

United States

Related Subject Headings

  • Ultrasonography, Interventional
  • Phantoms, Imaging
  • Humans
  • Gelatin
  • Equipment Design
  • Emergency Medicine
  • Emergency & Critical Care Medicine
  • Cost-Benefit Analysis
  • Catheterization, Peripheral
  • Catheterization, Central Venous
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Morrow, D. S., & Broder, J. (2015). Cost-effective, Reusable, Leak-resistant Ultrasound-guided Vascular Access Trainer. J Emerg Med, 49(3), 313–317. https://doi.org/10.1016/j.jemermed.2015.04.005
Morrow, Dustin S., and Joshua Broder. “Cost-effective, Reusable, Leak-resistant Ultrasound-guided Vascular Access Trainer.J Emerg Med 49, no. 3 (September 2015): 313–17. https://doi.org/10.1016/j.jemermed.2015.04.005.
Morrow DS, Broder J. Cost-effective, Reusable, Leak-resistant Ultrasound-guided Vascular Access Trainer. J Emerg Med. 2015 Sep;49(3):313–7.
Morrow, Dustin S., and Joshua Broder. “Cost-effective, Reusable, Leak-resistant Ultrasound-guided Vascular Access Trainer.J Emerg Med, vol. 49, no. 3, Sept. 2015, pp. 313–17. Pubmed, doi:10.1016/j.jemermed.2015.04.005.
Morrow DS, Broder J. Cost-effective, Reusable, Leak-resistant Ultrasound-guided Vascular Access Trainer. J Emerg Med. 2015 Sep;49(3):313–317.
Journal cover image

Published In

J Emerg Med

DOI

ISSN

0736-4679

Publication Date

September 2015

Volume

49

Issue

3

Start / End Page

313 / 317

Location

United States

Related Subject Headings

  • Ultrasonography, Interventional
  • Phantoms, Imaging
  • Humans
  • Gelatin
  • Equipment Design
  • Emergency Medicine
  • Emergency & Critical Care Medicine
  • Cost-Benefit Analysis
  • Catheterization, Peripheral
  • Catheterization, Central Venous