Inadvertent Transbronchial Insertion of Narrow-Bore Feeding Tubes Into the Pleural Space

Published

Journal Article

Four patients had soft, narrow-bore feeding tubes passed transbronchially into the pleural space. Guide wires were used to aid passage and may have contributed to this complication. The errant positioning of the tube was first noted roentgenographically in each case. Alternate methods of inserting such tubes in patients with altered airway and swallowing reflexes may be needed. Radiographic confirmation of the tube's position is essential before starting tube feedings. © 1984 American Medical Association. All rights reserved.