The Evolution of the Inflatable Penile Prosthesis Reservoir and Surgical Placement.

Published

Journal Article (Review)

The traditional inflatable penile prosthesis (IPP) reservoir placement is below the transversalis fascia in the space of Retzius. In 2002, Dr. Steve Wilson described ectopic reservoir placement, thereby providing a safe and effective alternative for implant surgeons. This new approach obviated the need for a second incision and decreased operative times during surgery. In the manuscript, he also described the introduction of a reservoir lock-out valve, which prevents autoinflation of the penile implant. The development of lockout valves and flat reservoirs has contributed to the early success and feasibility of submuscular placement techniques. Thirteen years after Dr. Wilson's pivotal study, this technique should be in the armamentarium of all urologic prosthetic surgeons. Accordingly, in certain subsets of patients, ectopic/ submuscular reservoir site placement should be considered a safe, effective alternative to standard reservoir placement in the space of Retzius.

Full Text

Duke Authors

Cited Authors

  • Hakky, T; Lentz, A; Sadeghi-Nejad, H; Khera, M

Published Date

  • November 2015

Published In

Volume / Issue

  • 12 Suppl 7 /

Start / End Page

  • 464 - 467

PubMed ID

  • 26565579

Pubmed Central ID

  • 26565579

Electronic International Standard Serial Number (EISSN)

  • 1743-6109

Digital Object Identifier (DOI)

  • 10.1111/jsm.13011

Language

  • eng

Conference Location

  • Netherlands