Skip to main content

Validity of a new computer-aided diagnosis imaging program to quantify nuclear cataract from slit-lamp photographs.

Publication ,  Journal Article
Cheung, CY-L; Li, H; Lamoureux, EL; Mitchell, P; Wang, JJ; Tan, AG; Johari, LK; Liu, J; Lim, JH; Aung, T; Wong, TY
Published in: Invest Ophthalmol Vis Sci
March 10, 2011

PURPOSE: To validate a new computer-aided diagnosis (CAD) imaging program for the assessment of nuclear lens opacity. METHODS: Slit-lamp lens photographs from the Singapore Malay Eye Study (SiMES) were graded using both the CAD imaging program and manual assessment method by a trained grader using the Wisconsin Cataract Grading System. Cataract was separately assessed clinically during the study using Lens Opacities Classification System III (LOCS III). The repeatability of CAD and Wisconsin grading methods were assessed using 160 paired images. The agreement between the CAD and Wisconsin grading methods, and the correlations of CAD with Wisconsin and LOCS III were assessed using the SiMES sample (5547 eyes from 2951 subjects). RESULTS: In assessing the repeatability, the coefficient of variation (CoV) was 8.10% (95% confidence interval [CI], 7.21-8.99), and the intraclass correlation coefficient (ICC) was 0.96 (95% CI, 0.93-0.96) for the CAD method. There was high agreement between the CAD and Wisconsin methods, with a mean difference (CAD minus Wisconsin) of -0.02 (95% limit of agreement, -0.91 and 0.87) and an ICC of 0.81 (95% CI, 0.80-0.82). CAD parameters were also significantly correlated with LOCS III grading (all P < 0.001). CONCLUSIONS: This new CAD imaging program assesses nuclear lens opacity with results comparable to the manual grading using the Wisconsin System. This study shows that an automated, precise, and quantitative assessment of nuclear cataract is possible.

Duke Scholars

Published In

Invest Ophthalmol Vis Sci

DOI

EISSN

1552-5783

Publication Date

March 10, 2011

Volume

52

Issue

3

Start / End Page

1314 / 1319

Location

United States

Related Subject Headings

  • Reproducibility of Results
  • Photography
  • Ophthalmology & Optometry
  • Middle Aged
  • Male
  • Lens Nucleus, Crystalline
  • Humans
  • Female
  • Diagnostic Techniques, Ophthalmological
  • Diagnostic Imaging
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Cheung, C.-L., Li, H., Lamoureux, E. L., Mitchell, P., Wang, J. J., Tan, A. G., … Wong, T. Y. (2011). Validity of a new computer-aided diagnosis imaging program to quantify nuclear cataract from slit-lamp photographs. Invest Ophthalmol Vis Sci, 52(3), 1314–1319. https://doi.org/10.1167/iovs.10-5427
Cheung, Carol Yim-lui, Huiqi Li, Ecosse L. Lamoureux, Paul Mitchell, Jie Jin Wang, Ava Grace Tan, Lily K. Johari, et al. “Validity of a new computer-aided diagnosis imaging program to quantify nuclear cataract from slit-lamp photographs.Invest Ophthalmol Vis Sci 52, no. 3 (March 10, 2011): 1314–19. https://doi.org/10.1167/iovs.10-5427.
Cheung CY-L, Li H, Lamoureux EL, Mitchell P, Wang JJ, Tan AG, et al. Validity of a new computer-aided diagnosis imaging program to quantify nuclear cataract from slit-lamp photographs. Invest Ophthalmol Vis Sci. 2011 Mar 10;52(3):1314–9.
Cheung, Carol Yim-lui, et al. “Validity of a new computer-aided diagnosis imaging program to quantify nuclear cataract from slit-lamp photographs.Invest Ophthalmol Vis Sci, vol. 52, no. 3, Mar. 2011, pp. 1314–19. Pubmed, doi:10.1167/iovs.10-5427.
Cheung CY-L, Li H, Lamoureux EL, Mitchell P, Wang JJ, Tan AG, Johari LK, Liu J, Lim JH, Aung T, Wong TY. Validity of a new computer-aided diagnosis imaging program to quantify nuclear cataract from slit-lamp photographs. Invest Ophthalmol Vis Sci. 2011 Mar 10;52(3):1314–1319.

Published In

Invest Ophthalmol Vis Sci

DOI

EISSN

1552-5783

Publication Date

March 10, 2011

Volume

52

Issue

3

Start / End Page

1314 / 1319

Location

United States

Related Subject Headings

  • Reproducibility of Results
  • Photography
  • Ophthalmology & Optometry
  • Middle Aged
  • Male
  • Lens Nucleus, Crystalline
  • Humans
  • Female
  • Diagnostic Techniques, Ophthalmological
  • Diagnostic Imaging