TH-A-9A-10: Prostate SBRT Delivery with Flattening-Filter-Free Mode: Benefit and Accuracy.

Publication ,  Journal Article
Li, T; Yuan, L; Sheng, Y; Wu, Q
Published in: Med Phys
June 2014

PURPOSE: Flattening-filter-free (FFF) beam mode offered on TrueBeam™ linac enables delivering IMRT at 2400 MU/min dose rate. This study investigates the benefit and delivery accuracy of using high dose rate in the context of prostate SBRT. METHODS: 8 prostate SBRT patients were retrospectively studied. In 5 cases treated with 600-MU/min dose rate, continuous prostate motion data acquired during radiation-beam-on was used to analyze motion range. In addition, the initial 1/3 of prostate motion trajectories during each radiation-beam-on was separated to simulate motion range if 2400-MU/min were used. To analyze delivery accuracy in FFF mode, MLC trajectory log files from an additional 3 cases treated at 2400-MU/min were acquired. These log files record MLC expected and actual positions every 20ms, and therefore can be used to reveal delivery accuracy. RESULTS: (1) Benefit. On average treatment at 600-MU/min takes 30s per beam; whereas 2400-MU/min requires only 11s. When shortening delivery time to ˜1/3, the prostate motion range was significantly smaller (p<0.001). Largest motion reduction occurred in Sup-Inf direction, from [-3.3mm, 2.1mm] to [-1.7mm, 1.7mm], followed by reduction from [-2.1mm, 2.4mm] to [-1.0mm, 2.4mm] in Ant-Pos direction. No change observed in LR direction [-0.8mm, 0.6mm]. The combined motion amplitude (vector norm) confirms that average motion and ranges are significantly smaller when beam-on was limited to the 1st 1/3 of actual delivery time. (2) Accuracy. Trajectory log file analysis showed excellent delivery accuracy with at 2400 MU/min. Most leaf deviations during beam-on were within 0.07mm (99-percentile). Maximum leaf-opening deviations during each beam-on were all under 0.1mm for all leaves. Dose-rate was maintained at 2400-MU/min during beam-on without dipping. CONCLUSION: Delivery prostate SBRT with 2400 MU/min is both beneficial and accurate. High dose rates significantly reduced both treatment time and intra-beam prostate motion range. Excellent delivery accuracy was confirmed with very small leaf motion deviation.

Duke Scholars

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

June 2014

Volume

41

Issue

6

Start / End Page

545

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Li, T., Yuan, L., Sheng, Y., & Wu, Q. (2014). TH-A-9A-10: Prostate SBRT Delivery with Flattening-Filter-Free Mode: Benefit and Accuracy. Med Phys, 41(6), 545. https://doi.org/10.1118/1.4889580
Li, T., L. Yuan, Y. Sheng, and Q. Wu. “TH-A-9A-10: Prostate SBRT Delivery with Flattening-Filter-Free Mode: Benefit and Accuracy.Med Phys 41, no. 6 (June 2014): 545. https://doi.org/10.1118/1.4889580.
Li, T., et al. “TH-A-9A-10: Prostate SBRT Delivery with Flattening-Filter-Free Mode: Benefit and Accuracy.Med Phys, vol. 41, no. 6, June 2014, p. 545. Pubmed, doi:10.1118/1.4889580.

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

June 2014

Volume

41

Issue

6

Start / End Page

545

Location

United States

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences