WE‐E‐201C‐03: Plate‐Specific Gain Map Correction for the Improvement of DQE in Computed Radiography

Publication ,  Conference
Schnell, E; Dobbins, J
Published in: Medical Physics
January 1, 2010

Purpose: To improve the NPS, and thus DQE, of CR images by correcting for pixel‐to‐pixel gain variations specific to each plate. Method and Materials: Ten high‐exposure, open field images were taken with an RQA5 spectrum, with a sixth generation CR plate suspended in air without a cassette. Image values were converted to exposure, the plates registered using fiducial dots on the plate, ten images averaged, and then high‐pass filtered to remove low frequency contributions from field inhomogeneity. A gain‐map was then produced by converting all pixel values in the average into fractions with mean of one. The resultant gain‐map of the plate was used to normalize subsequent single images to correct for pixel‐to‐pixel gain fluctuation. The normalized NPS (NNPS) for all images was calculated both with and without the gain‐map correction. Results: The NNPS with correction showed improvement over the non‐corrected case over the range of frequencies from 0.15 – 2.5 mm−1. At high exposure (40 mR), NNPS was 50–90% better with gain‐map correction than without. A small further improvement in NNPS was seen from careful registering of the gain‐map with subsequent images using small fiducial dots, because of slight misregistration during scanning. Conclusion: CR devices have not traditionally employed gain‐map corrections common with DR detectors because of the multiplicity of plates used with each reader. This study demonstrates that a simple gain‐map can be used to correct for the fixed‐pattern noise and thus improve the DQE of CR imaging. Such a method could easily be implemented by manufacturers because each plate has a unique bar code and the gain‐map could be stored for retrieval after plate reading. These experiments indicated that an improvement in NPS (and hence, DQE) is possible, depending on exposure level, over all frequencies with this technique. © 2010, American Association of Physicists in Medicine. All rights reserved.

Duke Scholars

Published In

Medical Physics

DOI

ISSN

0094-2405

Publication Date

January 1, 2010

Volume

37

Issue

6

Start / End Page

3439 / 3440

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Schnell, E., & Dobbins, J. (2010). WE‐E‐201C‐03: Plate‐Specific Gain Map Correction for the Improvement of DQE in Computed Radiography. In Medical Physics (Vol. 37, pp. 3439–3440). https://doi.org/10.1118/1.3469440
Schnell, E., and J. Dobbins. “WE‐E‐201C‐03: Plate‐Specific Gain Map Correction for the Improvement of DQE in Computed Radiography.” In Medical Physics, 37:3439–40, 2010. https://doi.org/10.1118/1.3469440.
Schnell, E., and J. Dobbins. “WE‐E‐201C‐03: Plate‐Specific Gain Map Correction for the Improvement of DQE in Computed Radiography.” Medical Physics, vol. 37, no. 6, 2010, pp. 3439–40. Scopus, doi:10.1118/1.3469440.

Published In

Medical Physics

DOI

ISSN

0094-2405

Publication Date

January 1, 2010

Volume

37

Issue

6

Start / End Page

3439 / 3440

Related Subject Headings

  • Nuclear Medicine & Medical Imaging
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences