SU‐GG‐I‐56: Testes and Uterine Dose Measurements from Scout View in Lumbar Puncture Procedure and Regular Pelvic CT Scan
Purpose: The purpose of this study was to measure the dose to testes and uterus from the scout views of lumbar puncture procedure and during a routine pelvic CT scan. Materials and methods: Five high sensitivity MOSFETs (TN‐1002RD, Best Medical, Canada) were placed in the testes (2 dosimeters), uterus (2 dosimeters) and skin (1 dosimeter in lower abdomen area) of an adult anthropomorphic phantom (Model 701‐D, CIRS, Norfolk, VA). A 16‐detector array Lightspeed GE CT scanner (GE Healthcare, Milwaukee, WI) was used. The phantom was placed in a prone position during the scout views for the lumbar puncture protocol. The scan parameters for anteroposterior (AP) scout were 120 kVp and 10 mA, and for the lateral (LAT) 120 kVp and 80 mA. The doses of the AP and LAT scout views were combined. During the regular pelvic CT scan the phantom was placed in supine position and scanned helically in automatic tube current modulation at 120 kVp and 440 mA (max tube current), the noise index was 11.57 and the CTDI
Duke Scholars
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Related Subject Headings
- Nuclear Medicine & Medical Imaging
- 5105 Medical and biological physics
- 4003 Biomedical engineering
- 1112 Oncology and Carcinogenesis
- 0903 Biomedical Engineering
- 0299 Other Physical Sciences
Citation
Published In
DOI
ISSN
Publication Date
Volume
Issue
Start / End Page
Related Subject Headings
- Nuclear Medicine & Medical Imaging
- 5105 Medical and biological physics
- 4003 Biomedical engineering
- 1112 Oncology and Carcinogenesis
- 0903 Biomedical Engineering
- 0299 Other Physical Sciences