Development of a 3D remote dosimetry protocol compatible with MRgIMRT.

Publication ,  Journal Article
Mein, S; Rankine, L; Adamovics, J; Li, H; Oldham, M
Published in: Med Phys
November 2017

PURPOSE: To develop a novel remote 3D dosimetry protocol to verify Magnetic Resonance-guided Radiation Therapy (MRgRT) treatments. The protocol was applied to investigate the accuracy of TG-119 IMRT irradiations delivered by the MRIdian® system (ViewRay® , Oakwood Village, OH, USA) allowing for a 48-hour delay between irradiation at a field institution and subsequent readout at a base institution. METHODS: The 3D dosimetry protocol utilizes a novel formulation of PRESAGE® radiochromic dosimeters developed for high postirradiation stability and compatibility with optical-CT readout. Optical-CT readout was performed with an in-house system utilizing telecentric lenses affording high-resolution scanning. The protocol was developed from preparatory experiments to characterize PRESAGE® response in relevant conditions. First, linearity and sensitivity of PRESAGE® dose-response in the presence of a magnetic field was evaluated in a small volume study (4 ml cuvettes) conducted under MRgRT conditions and irradiated with doses 0-15 Gy. Temporal and spatial stability of the dose-response were investigated in large volume studies utilizing large field-of-view (FOV) 2 kg cylindrical PRESAGE® dosimeters. Dosimeters were imaged at t = 1 hr and t = 48 hrs enabling the development of correction terms to model any observed spatial and temporal changes postirradiation. Polynomial correction factors for temporal and spatial changes in PRESAGE® dosimeters (CT and CR respectively) were obtained by numerical fitting to time-point data acquired in six irradiated dosimeters. A remote dosimetry protocol was developed where PRESAGE® change in optical-density (ΔOD) readings at time t = X (the irradiation to return shipment time interval) were corrected back to a convenient standard time t = 1 hr using the CT and CR corrections. This refined protocol was then applied to TG-119 (American Association of Physicists in Medicine, Task Group 119) plan deliveries on the MRIdian® system to evaluate the accuracy of MRgRT in these conditions. RESULTS: In the small volume study, in the presence of a 0.35 T magnetic field, PRESAGE® was observed to respond linearly (R2  = 0.9996) to Co-60 irradiation at t = 48 hrs postirradiation, within the dose ranges of 0 to 15 Gy, with a sensitivity of 0.0305(±0.003) ΔOD cm-1  Gy-1 . In the large volume studies, at t = 1 hr postirradiation, consistent linear response was observed, with average sensitivity of 0.0930 ± 0.002 ΔOD cm-1  Gy-1 . However, dosimeters gradually darkened with time (OD< 5% per day). A small radial dependence to the dosimeter sensitivity was measured (< 3% of maximum dose), which is attributed to a spherically symmetric dosimeter artifact arising from exothermic heating legacy in the PRESAGE® polyurethane substrate during curing. When applied to the TG-119 IMRT irradiations, the remote dosimetry protocol (including correction terms) yielded excellent line-profile and 3D gamma agreement for 3%/3 mm, 10% threshold (mean passing rate = 96.6% ± 4.0%). CONCLUSION: A novel 3D remote dosimetry protocol is introduced for validating off-site dosimetrically complex radiotherapy systems, including MRgRT. The protocol involves correcting for temporal and spatially dependent changes in PRESAGE® radiochromic dosimeters readout by optical-CT. Application of the protocol to TG-119 irradiations enabled verification of MRgRT dose distributions with high resolution.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

November 2017

Volume

44

Issue

11

Start / End Page

6018 / 6028

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Spatio-Temporal Analysis
  • Radiotherapy, Intensity-Modulated
  • Radiotherapy, Image-Guided
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Radiometry
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • 5105 Medical and biological physics
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Mein, S., Rankine, L., Adamovics, J., Li, H., & Oldham, M. (2017). Development of a 3D remote dosimetry protocol compatible with MRgIMRT. Med Phys, 44(11), 6018–6028. https://doi.org/10.1002/mp.12565
Mein, Stewart, Leith Rankine, John Adamovics, Harold Li, and Mark Oldham. “Development of a 3D remote dosimetry protocol compatible with MRgIMRT.Med Phys 44, no. 11 (November 2017): 6018–28. https://doi.org/10.1002/mp.12565.
Mein S, Rankine L, Adamovics J, Li H, Oldham M. Development of a 3D remote dosimetry protocol compatible with MRgIMRT. Med Phys. 2017 Nov;44(11):6018–28.
Mein, Stewart, et al. “Development of a 3D remote dosimetry protocol compatible with MRgIMRT.Med Phys, vol. 44, no. 11, Nov. 2017, pp. 6018–28. Pubmed, doi:10.1002/mp.12565.
Mein S, Rankine L, Adamovics J, Li H, Oldham M. Development of a 3D remote dosimetry protocol compatible with MRgIMRT. Med Phys. 2017 Nov;44(11):6018–6028.

Published In

Med Phys

DOI

EISSN

2473-4209

Publication Date

November 2017

Volume

44

Issue

11

Start / End Page

6018 / 6028

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Spatio-Temporal Analysis
  • Radiotherapy, Intensity-Modulated
  • Radiotherapy, Image-Guided
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Radiometry
  • Nuclear Medicine & Medical Imaging
  • Magnetic Resonance Imaging
  • 5105 Medical and biological physics