Monte Carlo calculation of imaging doses from diagnostic multidetector CT and kilovoltage cone-beam CT as part of prostate cancer treatment plans.

Publication ,  Journal Article
Ding, A; Gu, J; Trofimov, AV; Xu, XG
Published in: Medical physics
December 2010

To calculate imaging doses to the rectum, bladder, and femoral heads as part of a prostate cancer treatment plans, assuming an image guided radiation therapy (IGRT) procedure involving either the multidetector CT (MDCT) or kilovoltage cone-beam CT (kV CBCT).This study considered an IGRT treatment plan for a prostate carcinoma patient involving 50.4 Gy from 28 initial fractions and a boost of 28.8 Gy from 16 fractions. A total of 45 CT imaging procedures, each involving a MDCT or a kV CBCT scan procedure, were carefully modeled using the MCNPX code version 2.5.0. The MDCT scanner model is based on the GE LightSpeed 16-MDCT scanner and the kV CBCT scanner model is based on the Varian On-Board Imager using parameters reported by the CT manufacturers and literatures. A patient-specific treatment planning CT data set was used to construct the phantom for the dose calculation. The target, organs-at-risk (OARs), and background voxels in the CT data set were categorized into six tissue types according to CT numbers for Monte Carlo calculations.For a total of 45 imaging procedures, it was found that the rectum received 78.4 and 76.7 cGy from MDCT and kV CBCT, respectively. The bladder received slightly greater doses of 82.4 and 77.9 cGy, while the femoral heads received much higher doses of 182.3 and 141.3 cGy from MDCT and kV CBCT, respectively. To investigate the impact of these imaging doses on treatment planning, OAR doses from MDCT or kV CBCT imaging procedures were added to the corresponding dose matrix reported by the original treatment plans to construct dose volume histograms. It was found that after the imaging dose is added, the rectum volumes irradiated to 75 and 70 Gy increased from 13.9% and 21.2%, respectively, in the original plan to 14.8% and 21.8%. The bladder volumes receiving 80 Gy increased to 4.6% from 4.1% in the original plan and the volume receiving 75 Gy increased to 7.9% from 7.5%. All values remained within the tolerance levels: V70<25%, V75 <15% for rectum and V75 < 25%, V80 < 15% for bladder. The irradiation of femoral heads was also acceptable with no volume receiving >45 Gy.IGRT procedures can irradiate the OARs to an imaging dose level that is great enough to require careful evaluation and perhaps even adjustment of original treatment planning in order to still satisfy the dose constraints. This study only considered one patient CT because the CT x rays cover a relatively larger volume of the body and the dose distribution is considerably more uniform than those associated with the therapeutic beams. As a result, the dose to an organ from CT imaging doses does not vary much from one patient to the other for the same CT settings. One factor that would potentially affect such CT dose level is the size of the patient body. More studies are needed to develop accurate and convenient methods of accounting for the imaging doses as part of treatment planning.

Duke Scholars

Published In

Medical physics

DOI

EISSN

2473-4209

ISSN

0094-2405

Publication Date

December 2010

Volume

37

Issue

12

Start / End Page

6199 / 6204

Related Subject Headings

  • Radiotherapy, Intensity-Modulated
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Radiation Dosage
  • Prostatic Neoplasms
  • Organs at Risk
  • Nuclear Medicine & Medical Imaging
  • Monte Carlo Method
  • Male
  • Humans
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Ding, A., Gu, J., Trofimov, A. V., & Xu, X. G. (2010). Monte Carlo calculation of imaging doses from diagnostic multidetector CT and kilovoltage cone-beam CT as part of prostate cancer treatment plans. Medical Physics, 37(12), 6199–6204. https://doi.org/10.1118/1.3512791
Ding, Aiping, Jianwei Gu, Alexei V. Trofimov, and X George Xu. “Monte Carlo calculation of imaging doses from diagnostic multidetector CT and kilovoltage cone-beam CT as part of prostate cancer treatment plans.Medical Physics 37, no. 12 (December 2010): 6199–6204. https://doi.org/10.1118/1.3512791.
Ding, Aiping, et al. “Monte Carlo calculation of imaging doses from diagnostic multidetector CT and kilovoltage cone-beam CT as part of prostate cancer treatment plans.Medical Physics, vol. 37, no. 12, Dec. 2010, pp. 6199–204. Epmc, doi:10.1118/1.3512791.

Published In

Medical physics

DOI

EISSN

2473-4209

ISSN

0094-2405

Publication Date

December 2010

Volume

37

Issue

12

Start / End Page

6199 / 6204

Related Subject Headings

  • Radiotherapy, Intensity-Modulated
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Radiation Dosage
  • Prostatic Neoplasms
  • Organs at Risk
  • Nuclear Medicine & Medical Imaging
  • Monte Carlo Method
  • Male
  • Humans