Optimization and qualification of an Fc Array assay for assessments of antibodies against HIV-1/SIV.
Journal Article (Journal Article)
The Fc Array is a multiplexed assay that assesses the Fc domain characteristics of antigen-specific antibodies with the potential to evaluate up to 500 antigen specificities simultaneously. Antigen-specific antibodies are captured on antigen-conjugated beads and their functional capacity is probed via an array of Fc-binding proteins including antibody subclassing reagents, Fcγ receptors, complement proteins, and lectins. Here we present the results of the optimization and formal qualification of the Fc Array, performed in compliance with Good Clinical Laboratory Practice (GCLP) guidelines. Assay conditions were optimized for performance and reproducibility, and the final version of the assay was then evaluated for specificity, accuracy, precision, limits of detection and quantitation, linearity, range and robustness.
Full Text
Duke Authors
Cited Authors
- Brown, EP; Weiner, JA; Lin, S; Natarajan, H; Normandin, E; Barouch, DH; Alter, G; Sarzotti-Kelsoe, M; Ackerman, ME
Published Date
- April 2018
Published In
Volume / Issue
- 455 /
Start / End Page
- 24 - 33
PubMed ID
- 29395167
Pubmed Central ID
- PMC5851662
Electronic International Standard Serial Number (EISSN)
- 1872-7905
Digital Object Identifier (DOI)
- 10.1016/j.jim.2018.01.013
Language
- eng
Conference Location
- Netherlands