Skip to main content

Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph.

Publication ,  Journal Article
Vizzeri, G; Bowd, C; Medeiros, FA; Weinreb, RN; Zangwill, LM
Published in: J Glaucoma
August 2008

OBJECTIVE: Misalignment of the Stratus optical coherence tomograph scan circle placed by the operator around the optic nerve head (ONH) during each retinal nerve fiber layer (RNFL) examination can affect the instrument reproducibility and its theoretical ability to detect true structural changes in the RNFL thickness over time. We evaluated the effect of scan circle placement on RNFL measurements. DESIGN: Observational clinical study. METHODS: Sixteen eyes of 8 normal participants were examined using the Stratus optical coherence tomograph Fast RNFL thickness acquisition protocol (software version 4.0.7; Carl Zeiss Meditec, Dublin, CA). Four consecutive images were taken by the same operator with the circular scan centered on the optic nerve head. Four images each with the scan displaced superiorly, inferiorly, temporally, and nasally were also acquired. Differences in average and sectoral RNFL thicknesses were determined. For the centered scans, the coefficients of variation (CV) and the intraclass correlation coefficient for the average RNFL thickness measured were calculated. RESULTS: When the average RNFL thickness of the centered scans was compared with the average RNFL thickness of the displaced scans individually using analysis of variance with post-hoc analysis, no difference was found between the average RNFL thickness of the nasally (105.2 microm), superiorly (106.2 microm), or inferiorly (104.1 microm) displaced scans and the centered scans (106.4 microm). However, a significant difference (analysis of variance with Dunnett's test: F=8.82, P<0.0001) was found between temporally displaced scans (115.8 microm) and centered scans. Significant differences in sectoral RNFL thickness measurements were found between centered and each displaced scan. The coefficient of variation for average RNFL thickness was 1.75% and intraclass correlation coefficient was 0.95. CONCLUSIONS: In normal eyes, average RNFL thickness measurements are robust and similar with significant superior, inferior, and nasal scan displacement, but average RNFL thickness is greater when scans are displaced temporally. Parapapillary scan misalignment produces significant changes in RNFL assessment characterized by an increase in measured RNFL thickness in the quadrant in which the scan is closer to the disc, and a significant decrease in RNFL thickness in the quadrant in which the scan is displaced further from the optic disc.

Duke Scholars

Published In

J Glaucoma

DOI

EISSN

1536-481X

Publication Date

August 2008

Volume

17

Issue

5

Start / End Page

341 / 349

Location

United States

Related Subject Headings

  • Tomography, Optical Coherence
  • Retinal Ganglion Cells
  • Reproducibility of Results
  • Optic Disk
  • Ophthalmology & Optometry
  • Nerve Fibers
  • Male
  • Humans
  • Female
  • Diagnostic Errors
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Vizzeri, G., Bowd, C., Medeiros, F. A., Weinreb, R. N., & Zangwill, L. M. (2008). Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph. J Glaucoma, 17(5), 341–349. https://doi.org/10.1097/IJG.0b013e31815c3aeb
Vizzeri, Gianmarco, Christopher Bowd, Felipe A. Medeiros, Robert N. Weinreb, and Linda M. Zangwill. “Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph.J Glaucoma 17, no. 5 (August 2008): 341–49. https://doi.org/10.1097/IJG.0b013e31815c3aeb.
Vizzeri G, Bowd C, Medeiros FA, Weinreb RN, Zangwill LM. Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph. J Glaucoma. 2008 Aug;17(5):341–9.
Vizzeri, Gianmarco, et al. “Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph.J Glaucoma, vol. 17, no. 5, Aug. 2008, pp. 341–49. Pubmed, doi:10.1097/IJG.0b013e31815c3aeb.
Vizzeri G, Bowd C, Medeiros FA, Weinreb RN, Zangwill LM. Effect of improper scan alignment on retinal nerve fiber layer thickness measurements using Stratus optical coherence tomograph. J Glaucoma. 2008 Aug;17(5):341–349.

Published In

J Glaucoma

DOI

EISSN

1536-481X

Publication Date

August 2008

Volume

17

Issue

5

Start / End Page

341 / 349

Location

United States

Related Subject Headings

  • Tomography, Optical Coherence
  • Retinal Ganglion Cells
  • Reproducibility of Results
  • Optic Disk
  • Ophthalmology & Optometry
  • Nerve Fibers
  • Male
  • Humans
  • Female
  • Diagnostic Errors