Constancy checks of well-type ionization chambers with external-beam radiation units.

Journal Article (Journal Article)

Constancy checks of a well-type ionization chamber should be performed regularly as part of a quality assurance regime. The goal of this work was to test the feasibility of using a linear accelerator and an orthovoltage unit to check the constancy of a well-type chamber's response to an external radiation source. The reproducibility, linearity with dose, variation with dose-rate, and variation between energy-matched units of the well-type chamber response when exposed to 6 MV beams was examined. The robustness to errors in establishing the measurement conditions, including setting the source-to-surface distance and gantry angle, rotation of the chamber around the central axis of the beam, and the effect of changing the length of the chamber cable exposed to the field, were tested. The reproducibility and linearity with dose of the chamber response, and robustness to errors in establishing the measurement conditions for 100 kVp and 250 kVp beams from an orthovoltage unit, were also examined. The combined uncertainty, including contributions from errors in establishing the reference conditions, for well-type chamber measurements using a 6 MV beam from a linear accelerator is 1.0%. The combined uncertainties for measurements using 100 and 250 kVp beams were 1.8% and 1.5%, respectively. When focus-source distance errors were reduced to ≤ 1 mm, the combined uncertainties for the 100 and 250 kVp beams were 1.2% and 1.1%, respectively, when the dose to the chamber was confined to the linear region of the dose-response curve. The response of a well-type chamber should remain constant to within 1.2% when exposed to a constant dose from an external beam unit, if reference conditions can be reproducibly established. However, the uncertainty for establishing reference conditions for output measurements for an orthovoltage unit can be reduced, which would justify a reduction of the tolerance for constancy measurements.

Full Text

Duke Authors

Cited Authors

  • Hackett, SL; Davis, B; Nixon, A; Wyatt, R

Published Date

  • November 8, 2015

Published In

Volume / Issue

  • 16 / 6

Start / End Page

  • 508 - 514

PubMed ID

  • 26699571

Pubmed Central ID

  • PMC5691027

Electronic International Standard Serial Number (EISSN)

  • 1526-9914

Digital Object Identifier (DOI)

  • 10.1120/jacmp.v16i6.5608

Language

  • eng

Conference Location

  • United States