Machine learning for automated quality assurance in radiotherapy: A proof of principle using EPID data description.

Publication ,  Journal Article
El Naqa, I; Irrer, J; Ritter, TA; DeMarco, J; Al-Hallaq, H; Booth, J; Kim, G; Alkhatib, A; Popple, R; Perez, M; Farrey, K; Moran, JM
Published in: Medical physics
April 2019

Developing automated methods to identify task-driven quality assurance (QA) procedures is key toward increasing safety, efficacy, and efficiency. We investigate the use of machine learning (ML) methods for possible visualization, automation, and targeting of QA, and assess its performance using multi-institutional data.To enable automated analysis of QA data given its higher dimensional nature, we used nonlinear kernel mapping with support vector data description (SVDD) driven approaches. Instead of using labeled data as in typical support vector machine (SVM) applications, which requires exhaustive annotation, we applied a clustering extension of SVDD, which identifies the minimal enclosing hypersphere in the feature space defined by a kernel function separating normal operations from possible failures (i.e., outliers). In our case, QA test data are mapped by a Gaussian kernel to a higher dimensional feature space and then the minimal enclosing sphere was identified. This sphere, when mapped back to the input data space along the principal components, can separate the data into several components, each enclosing a separate cluster of QA points that could be used to evaluate tolerance boundaries and test reliability. We evaluated this approach for gantry sag, radiation field shift, and [multileaf collimator (MLC)] offset data acquired using electronic portal imaging devices (EPID), as representative examples.Data from eight LINACS and seven institutions (n = 119) were collected. A standardized EPID image of a phantom with fiducials provided deviation estimates between the radiation field and phantom center at four cardinal gantry angles. Deviation measurements in the horizontal direction (0°, 180°) were used to determine the gantry sag and deviations in the vertical direction (90°, 270°) were used to determine the field shift. These measurements were fed into the SVDD clustering algorithm with varying hypersphere radii (Gaussian widths). For gantry sag analysis, two clusters were identified one of which contained 2.5% of the outliers and also exceeded the 1 mm tolerance set by TG-142. In the case of field shifts, SVM clustering identified two distinct classes of measurements primarily driven by variations in the second principal component at 270°. Results from MLC analysis identified one outlier cluster (0.34%) along Leaf offset Constancy (LoC) axis that coincided with TG-142 limits.Machine learning methods based on SVDD clustering are promising for developing automated QA tools and providing insights into their reliability and reproducibility.

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Medical physics

DOI

EISSN

2473-4209

ISSN

0094-2405

Publication Date

April 2019

Volume

46

Issue

4

Start / End Page

1914 / 1921

Related Subject Headings

  • Radiotherapy Dosage
  • Quality Assurance, Health Care
  • Phantoms, Imaging
  • Particle Accelerators
  • Nuclear Medicine & Medical Imaging
  • Neoplasms
  • Machine Learning
  • Humans
  • Electrical Equipment and Supplies
  • Automation
 

Citation

APA
Chicago
ICMJE
MLA
NLM
El Naqa, I., Irrer, J., Ritter, T. A., DeMarco, J., Al-Hallaq, H., Booth, J., … Moran, J. M. (2019). Machine learning for automated quality assurance in radiotherapy: A proof of principle using EPID data description. Medical Physics, 46(4), 1914–1921. https://doi.org/10.1002/mp.13433
El Naqa, Issam, Jim Irrer, Tim A. Ritter, John DeMarco, Hania Al-Hallaq, Jeremy Booth, Grace Kim, et al. “Machine learning for automated quality assurance in radiotherapy: A proof of principle using EPID data description.Medical Physics 46, no. 4 (April 2019): 1914–21. https://doi.org/10.1002/mp.13433.
El Naqa I, Irrer J, Ritter TA, DeMarco J, Al-Hallaq H, Booth J, et al. Machine learning for automated quality assurance in radiotherapy: A proof of principle using EPID data description. Medical physics. 2019 Apr;46(4):1914–21.
El Naqa, Issam, et al. “Machine learning for automated quality assurance in radiotherapy: A proof of principle using EPID data description.Medical Physics, vol. 46, no. 4, Apr. 2019, pp. 1914–21. Epmc, doi:10.1002/mp.13433.
El Naqa I, Irrer J, Ritter TA, DeMarco J, Al-Hallaq H, Booth J, Kim G, Alkhatib A, Popple R, Perez M, Farrey K, Moran JM. Machine learning for automated quality assurance in radiotherapy: A proof of principle using EPID data description. Medical physics. 2019 Apr;46(4):1914–1921.

Published In

Medical physics

DOI

EISSN

2473-4209

ISSN

0094-2405

Publication Date

April 2019

Volume

46

Issue

4

Start / End Page

1914 / 1921

Related Subject Headings

  • Radiotherapy Dosage
  • Quality Assurance, Health Care
  • Phantoms, Imaging
  • Particle Accelerators
  • Nuclear Medicine & Medical Imaging
  • Neoplasms
  • Machine Learning
  • Humans
  • Electrical Equipment and Supplies