Determination of the neutron detection efficiency of an NE213 scintillator for En = 2.5 to 16 MeV using the 2 H(d, n)3 He reaction

Published

Journal Article

The absolute efficiency of an NE213 liquid scintillator of 12.7 cm diameter and 5.08 cm thickness was measured in the neutron energy range 2.5-16 MeV using the 2H(d, n)3He reaction as a source of monoenergetic neutrons. The efficiencies were measured at the time-of-flight facility of Triangle Universities Nuclear Laboratory (TUNL). The experimental data are compared to calculations from the Monte Carlo code NEFF of Physikalisch-Technische Bundesanstalt, Braunschweig, Germany (PTB).

Full Text

Duke Authors

Cited Authors

  • Al-Ohali, MA; Aksoy, A; Coban, A; Hanly, JM; Felsher, PD; Howell, CR; Tornow, W; Salinas, F; Walter, RL

Published Date

  • September 11, 1997

Published In

Volume / Issue

  • 396 / 3

Start / End Page

  • 388 - 393

International Standard Serial Number (ISSN)

  • 0168-9002

Digital Object Identifier (DOI)

  • 10.1016/S0168-9002(97)00701-8

Citation Source

  • Scopus