Response-function measurement of an NE213 scintillator using the 2 H(d, n)3 He reaction

Published

Journal Article

The response function of a 12.5 cm diameter NE213 scintillator detector has been measured over neutron energies ranging from 4.9 to 16.6 MeV. Beams of monoenergetic neutrons were produced using the 2H(d, n)3He reaction. The response function is about 12% lower than previously reported values for a similar detector obtained using a continuous spectrum of neutrons from a 252Cf fission source. However, the present response function agrees with portions of data sets obtained in published work that used monoenergetic neutron source reactions. © 1994.

Full Text

Duke Authors

Cited Authors

  • Aksoy, A; Coban, A; Naqvi, AA; Khiari, FZ; Hanly, JM; Howell, CR; Tornow, W; Felsher, PD; Al-Ohali, MA; Walter, RL

Published Date

  • January 1, 1994

Published In

Volume / Issue

  • 337 / 2-3

Start / End Page

  • 486 - 491

International Standard Serial Number (ISSN)

  • 0168-9002

Digital Object Identifier (DOI)

  • 10.1016/0168-9002(94)91118-5

Citation Source

  • Scopus