Half-cone beam collimation for triple-camera SPECT systems.
Journal Article (Journal Article)
UNLABELLED: Cone-beam collimators provide increased sensitivity at similar resolution compared to other collimators. The use of cone-beam collimators for brain imaging with triple-camera SPECT systems, however, results in truncation of the base of the brain because of clearance of the shoulders. A half-cone beam collimator does not have the problem of truncation. The objective of this study was to compare the performance characteristics of half-cone beam with parallel-beam and fan-beam collimators with similar resolution characteristics for SPECT imaging of the brain. METHODS: A half-cone beam collimator with the focal point located towards the base of the brain was built for a triple-camera SPECT system. Spatial resolutions and sensitivities of three collimators were measured. RESULTS: When 10-cm from the collimator surface, the planar spatial resolutions FWHM in mm (point source sensitivities in cps-MBq) for half-cone beam, fan-beam and parallel-beam collimators were 5.2 (85.6), 5.1 (55.6) and 5.9 (39.7), respectively. Image quality was evaluated using a three-dimensional Hoffman brain phantom and patient data. The deeper gray matter were more clearly visualized in the half-cone beam scans. CONCLUSION: Half-cone beam collimation provides higher sensitivity and offers the potential for improved brain imaging compared with parallel-beam and fan-beam collimation when used with a triple-camera SPECT system.
Full Text
Duke Authors
Cited Authors
- Li, J; Jaszczak, RJ; Van Mullekom, A; Scarfone, C; Greer, KL; Coleman, RE
Published Date
- March 1996
Published In
Volume / Issue
- 37 / 3
Start / End Page
- 498 - 502
PubMed ID
- 8772655
International Standard Serial Number (ISSN)
- 0161-5505
Language
- eng
Conference Location
- United States