Contrast echocardiography: influence of ultrasonic machine settings, mixing conditions, and pressurization on pixel intensity and microsphere size of Albunex solutions in vitro.

Publication ,  Journal Article
Ota, T; Hillman, ND; Craig, D; Kisslo, J; Smith, PK
Published in: J Am Soc Echocardiogr
1997

To use Albunex as a blood-flow tracer, the stability and consistency of microspheres under mixing conditions must be known. This study examined the effects of mixing conditions and machine settings on the size and echogenicity of Albunex solutions in vitro. Acoustic power, log compression, time-gain compensation, and transducer frequency were varied as Albunex solutions were imaged after mixing with magnetic stirring and pressurized. Higher acoustic power and lower transducer frequency decreased mean pixel intensity of Albunex solution images over time. Intensity, size, and number of Albunex microspheres were not significantly different between stirring speeds. The echogenicity of the Albunex solutions decreased with pressurization, and the critical pressure necessary to reduce the intensity to half its initial value increased with the logarithm of concentration (r = 0.91; p < 0.001). The microsphere size decreased with pressurization and remained smaller after pressure release (3.66 +/- 2.13 versus 1.47 +/- 0.95 microns; p < 0.01). These data indicate that acoustic power and transducer frequency may affect the physical properties of Albunex microspheres, decreasing mean videointensity. Pressure sensitivity of Albunex caused the decrease of videointensity and microsphere size.

Duke Scholars

Published In

J Am Soc Echocardiogr

DOI

ISSN

0894-7317

Publication Date

1997

Volume

10

Issue

1

Start / End Page

31 / 40

Location

United States

Related Subject Headings

  • Ultrasonics
  • Phantoms, Imaging
  • Microspheres
  • Echocardiography
  • Contrast Media
  • Cardiovascular System & Hematology
  • Albumins
  • 3201 Cardiovascular medicine and haematology
  • 1102 Cardiorespiratory Medicine and Haematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Ota, T., Hillman, N. D., Craig, D., Kisslo, J., & Smith, P. K. (1997). Contrast echocardiography: influence of ultrasonic machine settings, mixing conditions, and pressurization on pixel intensity and microsphere size of Albunex solutions in vitro. J Am Soc Echocardiogr, 10(1), 31–40. https://doi.org/10.1016/s0894-7317(97)80030-2
Ota, T., N. D. Hillman, D. Craig, J. Kisslo, and P. K. Smith. “Contrast echocardiography: influence of ultrasonic machine settings, mixing conditions, and pressurization on pixel intensity and microsphere size of Albunex solutions in vitro.J Am Soc Echocardiogr 10, no. 1 (1997): 31–40. https://doi.org/10.1016/s0894-7317(97)80030-2.
Ota, T., et al. “Contrast echocardiography: influence of ultrasonic machine settings, mixing conditions, and pressurization on pixel intensity and microsphere size of Albunex solutions in vitro.J Am Soc Echocardiogr, vol. 10, no. 1, 1997, pp. 31–40. Pubmed, doi:10.1016/s0894-7317(97)80030-2.
Journal cover image

Published In

J Am Soc Echocardiogr

DOI

ISSN

0894-7317

Publication Date

1997

Volume

10

Issue

1

Start / End Page

31 / 40

Location

United States

Related Subject Headings

  • Ultrasonics
  • Phantoms, Imaging
  • Microspheres
  • Echocardiography
  • Contrast Media
  • Cardiovascular System & Hematology
  • Albumins
  • 3201 Cardiovascular medicine and haematology
  • 1102 Cardiorespiratory Medicine and Haematology