Influence of epicardial patches on defibrillation threshold with nonthoracotomy lead configurations.

Publication ,  Journal Article
Fotuhi, PC; Ideker, RE; Idriss, SF; Callihan, RL; Walker, RG; Alt, EU
Published in: Circulation
November 15, 1995

BACKGROUND: In previous studies, epicardial patch electrodes decreased transthoracic defibrillation efficacy. We studied the effects of two inactive epicardial 14-cm2 titanium mesh patches on defibrillation energy requirements with nonthoracotomy internal lead configurations. METHODS AND RESULTS: A 6/6-millisecond biphasic shock wave-form was delivered via several electrode configurations 10 seconds after ventricular fibrillation was initiated with a 60-Hz generator. In two series, a total of 16 dogs (weight, 23.3 +/- 2.4 kg) underwent an up-down defibrillation protocol. In the first series, the defibrillation threshold (DFT) was determined for each electrode configuration in the presence of two inactive epicardial patches. In the second series, DFTs were determined in the presence of an inactive right ventricular (RV) or left ventricular (LV) patch alone. For several nonthoracotomy lead configurations tested in the first 8 dogs, the mean +/- SD DFT energy increased 49% to 97% with two inactive patches on the heart compared with no patches on the heart as follows: RV to superior vena caval (SVC) electrode, from 8.9 +/- 2.6 to 18.0 +/- 14.3 J; RV to SVC plus subcutaneous array electrode, from 7.0 +/- 2.4 to 10.7 +/- 5.3 J; RV to subcutaneous pectoral plate electrode, from 6.2 +/- 1.3 to 11.4 +/- 4.0 J (P < or = .05). The lowest DFT was achieved by defibrillating between the epicardial patches (3.8 +/- 3.3 J). The second series showed that DFT voltage requirements increased significantly for all three nonthoracotomy lead configurations with the inactive LV patch alone (P < or = .05) but not with the inactive RV patch alone. CONCLUSIONS: Inactive epicardial patches can significantly increase the defibrillation energy requirements for nonthoracotomy lead configurations. This negative impact may be due to an insulating effect of the patches and to a disturbance of the potential gradient field under the patches. If the same holds true in patients, these results have clinical implications. Functioning epicardial patch leads should be incorporated in the defibrillation lead system if already present. If the LV patch is nonfunctioning, such as because of a lead fracture, the marked increase in DFT due to an inactive LV patch calls for thorough DFT testing during surgery and, in selected patients, may necessitate patch removal to produce an effective transvenous-based system.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Circulation

DOI

ISSN

0009-7322

Publication Date

November 15, 1995

Volume

92

Issue

10

Start / End Page

3082 / 3088

Location

United States

Related Subject Headings

  • Ventricular Fibrillation
  • Titanium
  • Thoracotomy
  • Pericardium
  • Electrodes, Implanted
  • Electric Impedance
  • Electric Countershock
  • Dogs
  • Defibrillators, Implantable
  • Cardiovascular System & Hematology
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Fotuhi, P. C., Ideker, R. E., Idriss, S. F., Callihan, R. L., Walker, R. G., & Alt, E. U. (1995). Influence of epicardial patches on defibrillation threshold with nonthoracotomy lead configurations. Circulation, 92(10), 3082–3088. https://doi.org/10.1161/01.cir.92.10.3082
Fotuhi, P. C., R. E. Ideker, S. F. Idriss, R. L. Callihan, R. G. Walker, and E. U. Alt. “Influence of epicardial patches on defibrillation threshold with nonthoracotomy lead configurations.Circulation 92, no. 10 (November 15, 1995): 3082–88. https://doi.org/10.1161/01.cir.92.10.3082.
Fotuhi PC, Ideker RE, Idriss SF, Callihan RL, Walker RG, Alt EU. Influence of epicardial patches on defibrillation threshold with nonthoracotomy lead configurations. Circulation. 1995 Nov 15;92(10):3082–8.
Fotuhi, P. C., et al. “Influence of epicardial patches on defibrillation threshold with nonthoracotomy lead configurations.Circulation, vol. 92, no. 10, Nov. 1995, pp. 3082–88. Pubmed, doi:10.1161/01.cir.92.10.3082.
Fotuhi PC, Ideker RE, Idriss SF, Callihan RL, Walker RG, Alt EU. Influence of epicardial patches on defibrillation threshold with nonthoracotomy lead configurations. Circulation. 1995 Nov 15;92(10):3082–3088.

Published In

Circulation

DOI

ISSN

0009-7322

Publication Date

November 15, 1995

Volume

92

Issue

10

Start / End Page

3082 / 3088

Location

United States

Related Subject Headings

  • Ventricular Fibrillation
  • Titanium
  • Thoracotomy
  • Pericardium
  • Electrodes, Implanted
  • Electric Impedance
  • Electric Countershock
  • Dogs
  • Defibrillators, Implantable
  • Cardiovascular System & Hematology