Correction of geometric and refractive image distortions in optical coherence tomography applying Fermat's principle.

Published

Journal Article

We describe a methodology for quantitative image correction in OCT which includes procedures for correction of nonlinear axial scanning and non-telecentric scan patterns, as well as a novel approach for refraction correction in layered media based on Fermat's principle. The residual spatial error obtained in layered media with a fan-beam hand-held probe was reduced from several hundred micrometers to near the diffraction and coherence-length limits.

Full Text

Duke Authors

Cited Authors

  • Westphal, V; Rollins, A; Radhakrishnan, S; Izatt, J

Published Date

  • May 2002

Published In

Volume / Issue

  • 10 / 9

Start / End Page

  • 397 - 404

PubMed ID

  • 19436373

Pubmed Central ID

  • 19436373

Electronic International Standard Serial Number (EISSN)

  • 1094-4087

International Standard Serial Number (ISSN)

  • 1094-4087

Digital Object Identifier (DOI)

  • 10.1364/oe.10.000397

Language

  • eng