We present an analytical solution for the electrical potential and field established by a six-needle array electroporation electrode, which is used in vivo for cancer treatment and DNA delivery. The analytical solution closely matches the numerical solution obtained with the finite element method: the mean error is less than 0.6%.

Full Text

Duke Authors

Cited Authors

  • Dev, SB; Dhar, D; Krassowska, W

Published Date

  • November 2003

Published In

Volume / Issue

  • 50 / 11

Start / End Page

  • 1296 - 1300

PubMed ID

  • 14620000

Electronic International Standard Serial Number (EISSN)

  • 1558-2531

International Standard Serial Number (ISSN)

  • 0018-9294

Digital Object Identifier (DOI)

  • 10.1109/tbme.2003.818467