Electric field of a six-needle array electrode used in drug and DNA delivery in vivo: analytical versus numerical solution.
Journal Article (Journal Article)
We present an analytical solution for the electrical potential and field established by a six-needle array electroporation electrode, which is used in vivo for cancer treatment and DNA delivery. The analytical solution closely matches the numerical solution obtained with the finite element method: the mean error is less than 0.6%.
Full Text
Duke Authors
Cited Authors
- Dev, SB; Dhar, D; Krassowska, W
Published Date
- November 2003
Published In
Volume / Issue
- 50 / 11
Start / End Page
- 1296 - 1300
PubMed ID
- 14620000
Electronic International Standard Serial Number (EISSN)
- 1558-2531
International Standard Serial Number (ISSN)
- 0018-9294
Digital Object Identifier (DOI)
- 10.1109/tbme.2003.818467
Language