Cardiac ultrasound phantom using a porcine heart model.

Journal Article

A beating cardiac phantom has been developed using an excised porcine heart for use as a training tool in echocardiography. The heart is fixed with a formalin-based preservation method, housed in an optically transparent Lexan chamber, and undergoes hydraulic pumping to circulate a blood-mimicking fluid. The cardiac phantom has been used for a period of four months to produce images of excellent quality with ventricular wall motion typical of human subjects.

Full Text

Duke Authors

Cited Authors

  • Smith, SW; Lopath, PD; Adams, DB; Walcott, GP

Published Date

  • 1995

Published In

Volume / Issue

  • 21 / 5

Start / End Page

  • 693 - 697

PubMed ID

  • 8525559

International Standard Serial Number (ISSN)

  • 0301-5629

Language

  • eng

Conference Location

  • England