Angle independent ultrasonic blood flow detection by frame-to-frame correlation of B-mode images.

Journal Article (Journal Article)

We have previously reported initial clinical results of a novel blood velocity imaging technique utilizing a two-dimensional correlation search applied to consecutively acquired echoes. In this paper, we describe both the physical principles underlying this technique and test tank experiments which define its performance under a variety of conditions. The results indicate that, unlike Doppler flow imaging systems, this technique defines the flow velocity vector in two dimensions and is not subject to aliasing.

Full Text

Duke Authors

Cited Authors

  • Trahey, GE; Hubbard, SM; von Ramm, OT

Published Date

  • September 1988

Published In

Volume / Issue

  • 26 / 5

Start / End Page

  • 271 - 276

PubMed ID

  • 2970144

Electronic International Standard Serial Number (EISSN)

  • 1874-9968

International Standard Serial Number (ISSN)

  • 0041-624X

Digital Object Identifier (DOI)

  • 10.1016/0041-624x(88)90016-9

Language