SPECT quantification of technetium-99m microspheres within the canine lung.
Journal Article (Journal Article)
Attenuation compensated single photon emission CT (SPECT) count rates of 99mTc microsphere activities from 151 selected small regions of interest (ROIs) in an intact canine thorax were compared with the count rates obtained from scintillation camera images of similar ROIs from frozen slices of the thorax. A first-order method was used to compensate the SPECT and scintillation camera images for the effect of Compton scattering. The SPECT and scintillation camera count rates correlated well: r = 0.95 (RMS error 4.8 counts/s) for 19 X 19 X 13 mm thick ROIs. We conclude that regional quantification of 99mTc microspheres within the dog thorax is possible using camera-based SPECT systems.
Full Text
Duke Authors
Cited Authors
- Osborne, D; Jaszczak, RJ; Greer, K; Lischko, M; Coleman, RE
Published Date
- 1985
Published In
Volume / Issue
- 9 / 1
Start / End Page
- 73 - 77
PubMed ID
- 3871446
International Standard Serial Number (ISSN)
- 0363-8715
Digital Object Identifier (DOI)
- 10.1097/00004728-198501000-00015
Language
- eng
Conference Location
- United States