A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT.

Publication ,  Journal Article
Luo, JQ; Koral, KF; Ljungberg, M; Floyd, CE; Jaszczak, RJ
Published in: Phys Med Biol
January 1995

Using Monte Carlo simulation of 99Tcm single-photon-emission computed tomography (SPECT), we investigate the effects of tissue-background activity, tumour location, patient size, uncertainty of energy windows, and definition of tumour region on the accuracy of quantification. The dual-energy-window method of correction for Compton scattering is employed and the multiplier which yields correct activity for the VI as a whole calculated. The model is usually a sphere containing radioactive water located within a cylinder filled with a more dilute solution of radioactivity. Two simulation codes are employed. Reconstruction is by ML-EM algorithm with attenuation compensation. The scatter multiplier depends only slightly on the sphere location or the cylinder diameter. It also depends little on whether correction is before or after reconstruction. At low background level, it changes with VOI size, but not at higher background. For a geometrical VOI, it is 1.25 at zero background, decreases sharply to 0.56 for equal concentrations, and is 0.44 when the background concentration is very large. Quantification is accurate (less than 9% error) if the test background is reasonably close to that used in setting the universal scatter-multiplier value, or if the test backgrounds are always large and so is the universal-value background, but not if the test backgrounds cover a large range of values including zero. Results largely agree with those from experiment after the experimental data with background is re-evaluated with prejudice.

Duke Scholars

Published In

Phys Med Biol

DOI

ISSN

0031-9155

Publication Date

January 1995

Volume

40

Issue

1

Start / End Page

181 / 199

Location

England

Related Subject Headings

  • Tomography, Emission-Computed, Single-Photon
  • Technetium
  • Scattering, Radiation
  • Reproducibility of Results
  • Nuclear Medicine & Medical Imaging
  • Neoplasms
  • Monte Carlo Method
  • Models, Theoretical
  • Humans
  • 5105 Medical and biological physics
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Luo, J. Q., Koral, K. F., Ljungberg, M., Floyd, C. E., & Jaszczak, R. J. (1995). A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT. Phys Med Biol, 40(1), 181–199. https://doi.org/10.1088/0031-9155/40/1/015
Luo, J. Q., K. F. Koral, M. Ljungberg, C. E. Floyd, and R. J. Jaszczak. “A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT.Phys Med Biol 40, no. 1 (January 1995): 181–99. https://doi.org/10.1088/0031-9155/40/1/015.
Luo JQ, Koral KF, Ljungberg M, Floyd CE, Jaszczak RJ. A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT. Phys Med Biol. 1995 Jan;40(1):181–99.
Luo, J. Q., et al. “A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT.Phys Med Biol, vol. 40, no. 1, Jan. 1995, pp. 181–99. Pubmed, doi:10.1088/0031-9155/40/1/015.
Luo JQ, Koral KF, Ljungberg M, Floyd CE, Jaszczak RJ. A Monte Carlo investigation of dual-energy-window scatter correction for volume-of-interest quantification in 99Tcm SPECT. Phys Med Biol. 1995 Jan;40(1):181–199.
Journal cover image

Published In

Phys Med Biol

DOI

ISSN

0031-9155

Publication Date

January 1995

Volume

40

Issue

1

Start / End Page

181 / 199

Location

England

Related Subject Headings

  • Tomography, Emission-Computed, Single-Photon
  • Technetium
  • Scattering, Radiation
  • Reproducibility of Results
  • Nuclear Medicine & Medical Imaging
  • Neoplasms
  • Monte Carlo Method
  • Models, Theoretical
  • Humans
  • 5105 Medical and biological physics