Single photon emission computed tomography using multi-slice fan beam collimators

Published

Journal Article (Academic article)

Collimators have been built and evaluated for a SP/ECT system, consisting of two opposing LFOV scintillation cameras and a gantry, which have parallel collimation along the cranial-caudal axis of the patient and converging collimation in a direction perpendicular to this axis. A fan beam convolution algorithm reconstructs the transverse slices and provides partial compensation for γ-ray attenuation. It is concluded that the fan-beam collimator system, when compared to the conventional parallel-hole collimator system, produces simultaneous improvements in SP/ECT system sensitivity and resolution

Duke Authors

Cited Authors

  • Jaszczak, RJ; Chang, L-T; Murphy, PH

Published Date

  • 1979

Published In

  • Ieee Trans. Nucl. Sci. (Usa)

Volume / Issue

  • ns-26 / 1st1

Start / End Page

  • 610 - 618

Conference Location

  • Washington, DC, USA