Experimental Evaluation of a Tellurium-123m Transmission Source to Determine Attenuation Maps for SPECT1

Published

Journal Article

We propose to use a tellurium-123m line source to determine attenuation maps for reconstruction of technetium-99m images. To experimentally evaluate the performance of the transmission line source for the fast sequential approach for emission and transmission data acquisition, a prototype tellurium-123m line source was designed and manufactured. The line source has a length of 8 cm with total activity of 36–39 mCi at the time of the studies. Phantom and patient studies were conducted to evaluate the performance of the line source. The results of this study demonstrate that the tellurium-123m line source can be used to accurately determine attenuation maps for transmission studies with Tc-99m activity in the patient's myocardium. © 1995 IEEE.

Full Text

Duke Authors

Cited Authors

  • Wang, H; Jaszczak, RJ; McCormick, JW; Greer, KL; Coleman, RE

Published Date

  • January 1, 1995

Published In

Volume / Issue

  • 42 / 4

Start / End Page

  • 1214 - 1219

Electronic International Standard Serial Number (EISSN)

  • 1558-1578

International Standard Serial Number (ISSN)

  • 0018-9499

Digital Object Identifier (DOI)

  • 10.1109/23.467876

Citation Source

  • Scopus