A Comparison of PET Detector Modules Employing Rectangular and Round Photomultiplier Tubes

Publication ,  Journal Article
Cherry, SR; Tornai, MP; Levin, CS; Siegel, S; Hoffman, EJ; Andreaco, MS; Williams, CW
Published in: IEEE Transactions on Nuclear Science
January 1, 1995

We have compared the high resolution BGO detector blocks from the EXACT HR PET system which use two dual-cathode rectangular photomultiplier (PM) tubes with a new block design, the EXACT HR PLUS, which uses four round PM tubes. Despite the lower coupling area between photocathode and scintillator, the HR PLUS block compares favorably with the HR block. The energy resolution averages 20% for the HR PLUS block and 23% for the HR block, with efficiency variations of 17% in both blocks. Additional measurements were carried out on the HR PLUS block to characterize depth of interaction effects and cross-talk between elements. Coincidence line spread function measurements had a FWHM of 3.0 mm in the axial direction and 2.9 mm in the transaxial direction. In light of these results, limitations of the BGO block design are discussed and some solutions proposed. © 1995 IEEE.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

IEEE Transactions on Nuclear Science

DOI

EISSN

1558-1578

ISSN

0018-9499

Publication Date

January 1, 1995

Volume

42

Issue

4

Start / End Page

1064 / 1068

Related Subject Headings

  • Nuclear & Particles Physics
  • 5106 Nuclear and plasma physics
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences
  • 0202 Atomic, Molecular, Nuclear, Particle and Plasma Physics
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Cherry, S. R., Tornai, M. P., Levin, C. S., Siegel, S., Hoffman, E. J., Andreaco, M. S., & Williams, C. W. (1995). A Comparison of PET Detector Modules Employing Rectangular and Round Photomultiplier Tubes. IEEE Transactions on Nuclear Science, 42(4), 1064–1068. https://doi.org/10.1109/23.467748
Cherry, S. R., M. P. Tornai, C. S. Levin, S. Siegel, E. J. Hoffman, M. S. Andreaco, and C. W. Williams. “A Comparison of PET Detector Modules Employing Rectangular and Round Photomultiplier Tubes.” IEEE Transactions on Nuclear Science 42, no. 4 (January 1, 1995): 1064–68. https://doi.org/10.1109/23.467748.
Cherry SR, Tornai MP, Levin CS, Siegel S, Hoffman EJ, Andreaco MS, et al. A Comparison of PET Detector Modules Employing Rectangular and Round Photomultiplier Tubes. IEEE Transactions on Nuclear Science. 1995 Jan 1;42(4):1064–8.
Cherry, S. R., et al. “A Comparison of PET Detector Modules Employing Rectangular and Round Photomultiplier Tubes.” IEEE Transactions on Nuclear Science, vol. 42, no. 4, Jan. 1995, pp. 1064–68. Scopus, doi:10.1109/23.467748.
Cherry SR, Tornai MP, Levin CS, Siegel S, Hoffman EJ, Andreaco MS, Williams CW. A Comparison of PET Detector Modules Employing Rectangular and Round Photomultiplier Tubes. IEEE Transactions on Nuclear Science. 1995 Jan 1;42(4):1064–1068.

Published In

IEEE Transactions on Nuclear Science

DOI

EISSN

1558-1578

ISSN

0018-9499

Publication Date

January 1, 1995

Volume

42

Issue

4

Start / End Page

1064 / 1068

Related Subject Headings

  • Nuclear & Particles Physics
  • 5106 Nuclear and plasma physics
  • 0903 Biomedical Engineering
  • 0299 Other Physical Sciences
  • 0202 Atomic, Molecular, Nuclear, Particle and Plasma Physics