Mercuric iodide photodetector arrays for gamma-ray imaging

Published

Journal Article

We have developed a novel detector device, namely: a mercuric iodide photodetector array (HgI2 PDA) to be used with a scintillating crystal for X-ray and gamma-ray imaging. A HgI2 PDA prototype with 16-pixels has been evaluated with segmented scintillators (CsI(Tl)) coupled to the HgI2 PDA. Pixel energy resolution of 9.4% FWHM has been obtained at 122 keV for a 4 mm thick scintillator corresponding to 72% stopping at 140 keV, and pixel energy resolution of 11% FWHM has been obtained at 122 keV with the 10 mm thick scintillator.

Full Text

Duke Authors

Cited Authors

  • Patt, BE; Iwanczyk, JS; Wang, YJ; Tornai, MP; Levin, CS; Hoffman, EJ

Published Date

  • October 1, 1996

Published In

Volume / Issue

  • 380 / 1-2

Start / End Page

  • 295 - 300

International Standard Serial Number (ISSN)

  • 0168-9002

Digital Object Identifier (DOI)

  • 10.1016/S0168-9002(96)00492-5

Citation Source

  • Scopus