Pediatric cardiac-gated CT angiography: assessment of radiation dose.
Journal Article (Journal Article)
OBJECTIVE: The purpose of our study was to determine a dose range for cardiac-gated CT angiography (CTA) in children. MATERIALS AND METHODS: ECG-gated cardiac CTA simulating scanning of the heart was performed on an anthropomorphic phantom of a 5-year-old child on a 16-MDCT scanner using variable parameters (small field of view; 16 x 0.625 mm configuration; 0.5-second gantry cycle time; 0.275 pitch; 120 kVp at 110, 220, and 330 mA; and 80 kVp at 385 mA). Metal oxide semiconductor field effect transistor (MOSFET) technology measured 20 organ doses. Effective dose calculated using the dose-length product (DLP) was compared with effective dose determined from measured absorbed organ doses. RESULTS: Highest organ doses included breast (3.5-12.6 cGy), lung (3.3-12.1 cGy), and bone marrow (1.7-7.6 cGy). The 80 kVp/385 mA examination produced lower radiation doses to all organs than the 120 kVp/220 mA examination. MOSFET effective doses (+/- SD) were as follows: 110 mA: 7.4 mSv (+/- 0.6 mSv), 220 mA: 17.2 mSv (+/- 0.3 mSv), 330 mA: 25.7 mSv (+/- 0.3 mSv), 80 kVp/385 mA: 10.6 mSv (+/- 0.2 mSv). DLP effective doses for diagnostic runs were as follows: 110 mA: 8.7 mSv, 220 mA: 19 mSv, 330 mA: 28 mSv, 80 kVp/385 mA: 12 mSv. DLP effective doses exceeded MOSFET effective doses by 9.7-17.2%. CONCLUSION: Radiation doses for a 5-year-old during cardiac-gated CTA vary greatly depending on parameters. Organ doses can be high; the effective dose may reach 28.4 mSv. Further work, including determination of size-appropriate mA and image quality, is important before routine use of this technique in children.
Full Text
Duke Authors
Cited Authors
- Hollingsworth, CL; Yoshizumi, TT; Frush, DP; Chan, FP; Toncheva, G; Nguyen, G; Lowry, CR; Hurwitz, LM
Published Date
- July 2007
Published In
Volume / Issue
- 189 / 1
Start / End Page
- 12 - 18
PubMed ID
- 17579144
Electronic International Standard Serial Number (EISSN)
- 1546-3141
Digital Object Identifier (DOI)
- 10.2214/AJR.06.1507
Language
- eng
Conference Location
- United States