MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children.

Published

Journal Article

OBJECTIVE: The purpose of our study was to measure radiation dose to the orbit during pediatric cranial CT with and without bismuth shielding using a novel dosimetry system. Cranial CT was performed on a pediatric anthropomorphic phantom, with and without bismuth eye shields. A solid-state metal oxide semiconductor field effect transistor (MOSFET) dosimeter was used to obtain real-time dose measurements. CONCLUSION: Bismuth shielding reduced radiation dose to the eye by up to 42%; shield artifact fell outside the diagnostic area of interest.

Full Text

Duke Authors

Cited Authors

  • Mukundan, S; Wang, PI; Frush, DP; Yoshizumi, T; Marcus, J; Kloeblen, E; Moore, M

Published Date

  • June 2007

Published In

Volume / Issue

  • 188 / 6

Start / End Page

  • 1648 - 1650

PubMed ID

  • 17515389

Pubmed Central ID

  • 17515389

Electronic International Standard Serial Number (EISSN)

  • 1546-3141

Digital Object Identifier (DOI)

  • 10.2214/AJR.06.1146

Language

  • eng

Conference Location

  • United States