Measurement of vertebral cortical integrity during pedicle exploration for intrapedicular fixation.

Published

Journal Article

STUDY DESIGN: This study determined the predictive ability of electrical impedance measurement in detecting cortical perforation in a porcine model of pedicular exploration. OBJECTIVE: This study tested the hypothesis that a large decrease in electrical impedance would occur as a result of perforation of the vertebral cortex by the pedicle probe. SUMMARY OF BACKGROUND DATA: The resistivity of cortical bone has been reported to be 25 to 100 times greater than that of soft tissues. METHODS: A total of 42 pedicles of the lumbar spines of six swine were explored using the instrumented pedicle probes. RESULTS: Using a 1 microAmp 30-Hz current source, measurement of electrical impedance predicted cortical rupture with a sensitivity, specificity, and accuracy of 95%. Maximum applied voltages of 2.8 mV did not result in myogenic stimulus. CONCLUSIONS: Electrical impedance measurement provides an accurate real-time measurement of cortical perforation. This technique is adapted readily for use with pedicular screws and screw tape. Further investigation to determine the clinical use of this technique is recommended.