Accuracy of ROPtool vs individual examiners in assessing retinal vascular tortuosity.

Journal Article

OBJECTIVE: To prospectively determine if tortuosity assessment by a computer program (ROPtool) that traces retinal blood vessels and measures their tortuosity was more accurate than that of individual pediatric ophthalmologists. METHODS: One hundred eighty-five high-quality RetCam images from premature infants were circulated to 3 retinopathy of prematurity (ROP) experts and 3 other pediatric ophthalmologists ("examiners") who graded the tortuosity in each quadrant as normal, pre-plus, or plus. These same images were analyzed using ROPtool. RESULTS: Using expert consensus as the standard, ROPtool's overall accuracy of 95% (175 of 185) for identifying tortuosity sufficient for plus disease was similar to that of examiner 1 (93%; 172 of 185; P = .50), examiner 2 (93%; 172 of 185; P = .50), and examiner 3 (91%; 168 of 185; P = .10). ROPtool's sensitivity of 97% (36 of 37) compared favorably with that of examiner 1 (65%; 24 of 37; P < .001), examiner 2 (70%; 26 of 37; P < .001), and examiner 3 (81%; 30 of 37; P = .06). CONCLUSION: Computer-assisted analysis of retinal images can potentially reduce subjectivity in the diagnosis of plus disease and optimize timing of follow-up and treatment for ROP.

Full Text

Duke Authors

Cited Authors

  • Wallace, DK; Freedman, SF; Zhao, Z; Jung, S-H

Published Date

  • November 2007

Published In

Volume / Issue

  • 125 / 11

Start / End Page

  • 1523 - 1530

PubMed ID

  • 17998514

International Standard Serial Number (ISSN)

  • 0003-9950

Digital Object Identifier (DOI)

  • 10.1001/archopht.125.11.1523

Language

  • eng

Conference Location