Quantitative evaluation of cone-beam computed tomography in target volume definition for offline image-guided radiation therapy of prostate cancer.

Published

Journal Article

To quantitatively evaluate cone-beam CT (CBCT) in target volume definition in an offline image guidance environment.Fifteen patients each with five helical CTs (HCT) and eight CBCTs were included. A single physician manually delineated prostate and seminal vesicles (SVs) on each CT. The clinical target volume (CTV) was prostate for low risk group (G1), plus SVs for intermediate risk group (G2). The internal target volumes (ITVs) on CBCT (ITV(CBCT)) were constructed and compared with ITV(HCT). The following comparisons were performed: CTV and ITV in HCT and CBCT; similarity of ITVs using overlap index (OI); surface differences between ITVs; quality assurance of ITV(CBCT) using CTV from weekly CBCT; and dosimetric evaluations of ITV(HCT) coverage on plans from ITV(CBCT).There was no statistical significant difference of CTV or ITV. The ITV OIs were 91%/88% for G1/G2 patients. They improved significantly with 1-2mm margins. Therefore, the ITVs were mostly within 2mm. The CTVs from weekly CBCT had >95% overlap with ITV(CBCT). The ITV dose differences (D(95), and D(mean)) were <0.3%.It is feasible to use CBCT for target definition in offline image guidance, thereby eliminating the separate helical CT scan process.

Full Text

Duke Authors

Cited Authors

  • Wang, W; Wu, Q; Yan, D

Published Date

  • January 2010

Published In

Volume / Issue

  • 94 / 1

Start / End Page

  • 71 - 75

PubMed ID

  • 19897268

Pubmed Central ID

  • 19897268

Electronic International Standard Serial Number (EISSN)

  • 1879-0887

International Standard Serial Number (ISSN)

  • 0167-8140

Digital Object Identifier (DOI)

  • 10.1016/j.radonc.2009.10.005

Language

  • eng