Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans.

Publication ,  Journal Article
Siebers, JV; Keall, PJ; Wu, Q; Williamson, JF; Schmidt-Ullrich, RK
Published in: Int J Radiat Oncol Biol Phys
October 1, 2005

PURPOSE: The purpose of this study is to determine dose delivery errors that could result from random and systematic setup errors for head-and-neck patients treated using the simultaneous integrated boost (SIB)-intensity-modulated radiation therapy (IMRT) technique. METHODS AND MATERIALS: Twenty-four patients who participated in an intramural Phase I/II parotid-sparing IMRT dose-escalation protocol using the SIB treatment technique had their dose distributions reevaluated to assess the impact of random and systematic setup errors. The dosimetric effect of random setup error was simulated by convolving the two-dimensional fluence distribution of each beam with the random setup error probability density distribution. Random setup errors of sigma = 1, 3, and 5 mm were simulated. Systematic setup errors were simulated by randomly shifting the patient isocenter along each of the three Cartesian axes, with each shift selected from a normal distribution. Systematic setup error distributions with Sigma = 1.5 and 3.0 mm along each axis were simulated. Combined systematic and random setup errors were simulated for sigma = Sigma = 1.5 and 3.0 mm along each axis. For each dose calculation, the gross tumor volume (GTV) received by 98% of the volume (D(98)), clinical target volume (CTV) D(90), nodes D(90), cord D(2), and parotid D(50) and parotid mean dose were evaluated with respect to the plan used for treatment for the structure dose and for an effective planning target volume (PTV) with a 3-mm margin. RESULTS: Simultaneous integrated boost-IMRT head-and-neck treatment plans were found to be less sensitive to random setup errors than to systematic setup errors. For random-only errors, errors exceeded 3% only when the random setup error sigma exceeded 3 mm. Simulated systematic setup errors with Sigma = 1.5 mm resulted in approximately 10% of plan having more than a 3% dose error, whereas a Sigma = 3.0 mm resulted in half of the plans having more than a 3% dose error and 28% with a 5% dose error. Combined random and systematic dose errors with sigma = Sigma = 3.0 mm resulted in more than 50% of plans having at least a 3% dose error and 38% of the plans having at least a 5% dose error. Evaluation with respect to a 3-mm expanded PTV reduced the observed dose deviations greater than 5% for the sigma = Sigma = 3.0 mm simulations to 5.4% of the plans simulated. CONCLUSIONS: Head-and-neck SIB-IMRT dosimetric accuracy would benefit from methods to reduce patient systematic setup errors. When GTV, CTV, or nodal volumes are used for dose evaluation, plans simulated including the effects of random and systematic errors deviate substantially from the nominal plan. The use of PTVs for dose evaluation in the nominal plan improves agreement with evaluated GTV, CTV, and nodal dose values under simulated setup errors. PTV concepts should be used for SIB-IMRT head-and-neck squamous cell carcinoma patients, although the size of the margins may be less than those used with three-dimensional conformal radiation therapy.

Duke Scholars

Published In

Int J Radiat Oncol Biol Phys

DOI

ISSN

0360-3016

Publication Date

October 1, 2005

Volume

63

Issue

2

Start / End Page

422 / 433

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Radiotherapy, Conformal
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Oncology & Carcinogenesis
  • Humans
  • Head and Neck Neoplasms
  • Algorithms
  • 5105 Medical and biological physics
  • 3407 Theoretical and computational chemistry
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Siebers, J. V., Keall, P. J., Wu, Q., Williamson, J. F., & Schmidt-Ullrich, R. K. (2005). Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans. Int J Radiat Oncol Biol Phys, 63(2), 422–433. https://doi.org/10.1016/j.ijrobp.2005.02.029
Siebers, Jeffrey V., Paul J. Keall, Qiuwen Wu, Jeffrey F. Williamson, and Rupert K. Schmidt-Ullrich. “Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans.Int J Radiat Oncol Biol Phys 63, no. 2 (October 1, 2005): 422–33. https://doi.org/10.1016/j.ijrobp.2005.02.029.
Siebers JV, Keall PJ, Wu Q, Williamson JF, Schmidt-Ullrich RK. Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans. Int J Radiat Oncol Biol Phys. 2005 Oct 1;63(2):422–33.
Siebers, Jeffrey V., et al. “Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans.Int J Radiat Oncol Biol Phys, vol. 63, no. 2, Oct. 2005, pp. 422–33. Pubmed, doi:10.1016/j.ijrobp.2005.02.029.
Siebers JV, Keall PJ, Wu Q, Williamson JF, Schmidt-Ullrich RK. Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans. Int J Radiat Oncol Biol Phys. 2005 Oct 1;63(2):422–433.
Journal cover image

Published In

Int J Radiat Oncol Biol Phys

DOI

ISSN

0360-3016

Publication Date

October 1, 2005

Volume

63

Issue

2

Start / End Page

422 / 433

Location

United States

Related Subject Headings

  • Sensitivity and Specificity
  • Radiotherapy, Conformal
  • Radiotherapy Planning, Computer-Assisted
  • Radiotherapy Dosage
  • Oncology & Carcinogenesis
  • Humans
  • Head and Neck Neoplasms
  • Algorithms
  • 5105 Medical and biological physics
  • 3407 Theoretical and computational chemistry