Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study.

Publication ,  Journal Article
Schindera, ST; Diedrichsen, L; Müller, HC; Rusch, O; Marin, D; Schmidt, B; Raupach, R; Vock, P; Szucs-Farkas, Z
Published in: Radiology
August 2011

PURPOSE: To assess the diagnostic accuracy, image quality, and radiation dose of an iterative reconstruction algorithm compared with a filtered back projection (FBP) algorithm for abdominal computed tomography (CT) at different tube voltages. MATERIALS AND METHODS: A custom liver phantom with 45 simulated hypovascular liver tumors (diameters of 5, 10, and 15 mm; tumor-to-liver contrast of 10, 25, and 50 HU) was placed in a cylindrical water container that mimicked an intermediate-sized patient. The phantom was scanned at 120, 100, and 80 kVp. The CT data sets were reconstructed with FBP and iterative reconstruction. The image noise was measured, and the contrast-to-noise ratio (CNR) of the tumors was calculated. The radiation dose was assessed with the volume CT dose index. Tumor detection was independently performed by three radiologists. Statistical analysis included analysis of variance. RESULTS: Compared with the FBP data set at 120 kVp, the iterative reconstruction data set collected at 100 kVp demonstrated significantly lower mean image noise (20.9 and 16.7 HU, respectively; P < .001) and greater mean CNRs for the simulated tumors (P < .001). The iterative reconstruction data set collected at 120 kVp yielded the highest sensitivity for tumor detection, while the FBP data set at 80 kVp yielded the lowest. The sensitivity for the iterative reconstruction data set at 100 kVp was comparable with that for the FBP data set at 120 kVp (79.3% and 74.9%, respectively; P > .99). The volume CT dose index decreased by 39.8% between the 120-kVp protocol and the 100-kVp protocol and by 70.3% between the 120-kVp protocol and the 80-kVp protocol. CONCLUSION: Results of this phantom study suggest that a 100-kVp abdominal CT protocol with an iterative reconstruction algorithm for simulated intermediate-sized patients increases the image quality and maintains the diagnostic accuracy at a reduced radiation dose when compared with a 120-kVp protocol with an FBP algorithm.

Duke Scholars

Published In

Radiology

DOI

EISSN

1527-1315

Publication Date

August 2011

Volume

260

Issue

2

Start / End Page

454 / 462

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Radiography, Abdominal
  • Radiographic Image Interpretation, Computer-Assisted
  • Radiation Dosage
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Liver Neoplasms
  • Humans
  • Analysis of Variance
  • Algorithms
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Schindera, S. T., Diedrichsen, L., Müller, H. C., Rusch, O., Marin, D., Schmidt, B., … Szucs-Farkas, Z. (2011). Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study. Radiology, 260(2), 454–462. https://doi.org/10.1148/radiol.11102217
Schindera, Sebastian T., Lars Diedrichsen, Hubert C. Müller, Oliver Rusch, Daniele Marin, Bernhard Schmidt, Rainer Raupach, Peter Vock, and Zsolt Szucs-Farkas. “Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study.Radiology 260, no. 2 (August 2011): 454–62. https://doi.org/10.1148/radiol.11102217.
Schindera, Sebastian T., et al. “Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study.Radiology, vol. 260, no. 2, Aug. 2011, pp. 454–62. Pubmed, doi:10.1148/radiol.11102217.
Schindera ST, Diedrichsen L, Müller HC, Rusch O, Marin D, Schmidt B, Raupach R, Vock P, Szucs-Farkas Z. Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of diagnostic accuracy, image quality, and radiation dose in a phantom study. Radiology. 2011 Aug;260(2):454–462.

Published In

Radiology

DOI

EISSN

1527-1315

Publication Date

August 2011

Volume

260

Issue

2

Start / End Page

454 / 462

Location

United States

Related Subject Headings

  • Tomography, X-Ray Computed
  • Radiography, Abdominal
  • Radiographic Image Interpretation, Computer-Assisted
  • Radiation Dosage
  • Phantoms, Imaging
  • Nuclear Medicine & Medical Imaging
  • Liver Neoplasms
  • Humans
  • Analysis of Variance
  • Algorithms