Accuracy of registration of PET, SPECT and MR images of a brain phantom.

Publication ,  Journal Article
Turkington, TG; Jaszczak, RJ; Pelizzari, CA; Harris, CC; MacFall, JR; Hoffman, JM; Coleman, RE
Published in: J Nucl Med
September 1993

Accuracy of a surface-fitting algorithm for three-dimensional image registration of single photon emission computed tomography (SPECT), positron emission tomography (PET), and magnetic resonance (MR) images was tested using a three-dimensional, water-fillable brain phantom. Multislice or volume image sets were acquired for each modality. Small fiducial markers were attached to assess accuracy of surface fitting and provide an alternate fitting technique. A maximum gradient technique was found to work well for SPECT and PET edge detection. Transformation parameters for translation, rotation and scaling were determined by surface fit to match each SPECT and PET scan with MR images. Using the markers, overall translation errors were found to be < 2 mm in each direction and rotational errors < 2 degrees in every case. Errors for specific internal regions were also determined to be < 2 mm for most regions, with only a few fits resulting in errors > 3 mm for some cortical regions. Results indicate surface fitting to be sufficiently accurate for visual comparison of registered images and for enhanced SPECT and PET region of interest (ROI) determination and image reconstruction.

Duke Scholars

Published In

J Nucl Med

ISSN

0161-5505

Publication Date

September 1993

Volume

34

Issue

9

Start / End Page

1587 / 1594

Location

United States

Related Subject Headings

  • Tomography, Emission-Computed, Single-Photon
  • Tomography, Emission-Computed
  • Nuclear Medicine & Medical Imaging
  • Models, Structural
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Brain
  • Algorithms
  • 3202 Clinical sciences
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Turkington, T. G., Jaszczak, R. J., Pelizzari, C. A., Harris, C. C., MacFall, J. R., Hoffman, J. M., & Coleman, R. E. (1993). Accuracy of registration of PET, SPECT and MR images of a brain phantom. J Nucl Med, 34(9), 1587–1594.
Turkington, T. G., R. J. Jaszczak, C. A. Pelizzari, C. C. Harris, J. R. MacFall, J. M. Hoffman, and R. E. Coleman. “Accuracy of registration of PET, SPECT and MR images of a brain phantom.J Nucl Med 34, no. 9 (September 1993): 1587–94.
Turkington TG, Jaszczak RJ, Pelizzari CA, Harris CC, MacFall JR, Hoffman JM, et al. Accuracy of registration of PET, SPECT and MR images of a brain phantom. J Nucl Med. 1993 Sep;34(9):1587–94.
Turkington, T. G., et al. “Accuracy of registration of PET, SPECT and MR images of a brain phantom.J Nucl Med, vol. 34, no. 9, Sept. 1993, pp. 1587–94.
Turkington TG, Jaszczak RJ, Pelizzari CA, Harris CC, MacFall JR, Hoffman JM, Coleman RE. Accuracy of registration of PET, SPECT and MR images of a brain phantom. J Nucl Med. 1993 Sep;34(9):1587–1594.

Published In

J Nucl Med

ISSN

0161-5505

Publication Date

September 1993

Volume

34

Issue

9

Start / End Page

1587 / 1594

Location

United States

Related Subject Headings

  • Tomography, Emission-Computed, Single-Photon
  • Tomography, Emission-Computed
  • Nuclear Medicine & Medical Imaging
  • Models, Structural
  • Magnetic Resonance Imaging
  • Image Processing, Computer-Assisted
  • Humans
  • Brain
  • Algorithms
  • 3202 Clinical sciences