Comparison of imaging properties of a computed radiography system and screen-film systems.

Publication ,  Journal Article
Sanada, S; Doi, K; Xu, XW; Yin, FF; Giger, ML; MacMahon, H
Published in: Med Phys
1991

To compare the diagnostic quality of images obtained with a computed radiography (CR) system based on storage phosphor technology with that obtained with conventional screen-film systems, a dual-image recording technique was devised. With this technique, a CR imaging plate is placed behind a screen-film system in a conventional cassette. This makes it possible to obtain two images simultaneously, one from each system, in a clinical examination with the same patient positioning, the same degree of patient motion, the same geometric unsharpness, and no additional exposure. The modulation transfer functions (MTFs) of the CR system with and without the dual-image recording technique were greater at low frequencies, but lower at high frequencies, that the MTFs of the screen-film systems used. The noise Wiener spectra of the CR images at the plane of the imaging plate were greater than those of the screen-film systems, but were comparable to those of the screen-film systems at the plane of the printed film due to the reduction in image size. Clinical chest images obtained with the dual-image recording technique appeared comparable, probably because of the image size reduction and the use of mild unsharp mask processing.

Duke Scholars

Altmetric Attention Stats
Dimensions Citation Stats

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

1991

Volume

18

Issue

3

Start / End Page

414 / 420

Location

United States

Related Subject Headings

  • X-Ray Intensifying Screens
  • Radiography, Thoracic
  • Radiographic Image Enhancement
  • Nuclear Medicine & Medical Imaging
  • Luminescent Measurements
  • Humans
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Sanada, S., Doi, K., Xu, X. W., Yin, F. F., Giger, M. L., & MacMahon, H. (1991). Comparison of imaging properties of a computed radiography system and screen-film systems. Med Phys, 18(3), 414–420. https://doi.org/10.1118/1.596688
Sanada, S., K. Doi, X. W. Xu, F. F. Yin, M. L. Giger, and H. MacMahon. “Comparison of imaging properties of a computed radiography system and screen-film systems.Med Phys 18, no. 3 (1991): 414–20. https://doi.org/10.1118/1.596688.
Sanada S, Doi K, Xu XW, Yin FF, Giger ML, MacMahon H. Comparison of imaging properties of a computed radiography system and screen-film systems. Med Phys. 1991;18(3):414–20.
Sanada, S., et al. “Comparison of imaging properties of a computed radiography system and screen-film systems.Med Phys, vol. 18, no. 3, 1991, pp. 414–20. Pubmed, doi:10.1118/1.596688.
Sanada S, Doi K, Xu XW, Yin FF, Giger ML, MacMahon H. Comparison of imaging properties of a computed radiography system and screen-film systems. Med Phys. 1991;18(3):414–420.

Published In

Med Phys

DOI

ISSN

0094-2405

Publication Date

1991

Volume

18

Issue

3

Start / End Page

414 / 420

Location

United States

Related Subject Headings

  • X-Ray Intensifying Screens
  • Radiography, Thoracic
  • Radiographic Image Enhancement
  • Nuclear Medicine & Medical Imaging
  • Luminescent Measurements
  • Humans
  • 5105 Medical and biological physics
  • 4003 Biomedical engineering
  • 1112 Oncology and Carcinogenesis
  • 0903 Biomedical Engineering