Evaluation of edge-induced streaking artifacts in CT scanners.

Journal Article

A phantom is presented which permits the evaluation of streaking artifacts produced in CT reconstructions by abrupt edges. Its application is demonstrated by results obtained from nine CT scanners. It is observed that even in regions where streaking is not readily apparent, edge-induced artifacts can significantly increase the variance in the reconstruction.

Full Text

Duke Authors

Cited Authors

  • Sheridan, WT; Keller, MR; O'Connor, CM; Brooks, RA; Hanson, KM

Published Date

  • March 1, 1980

Published In

Volume / Issue

  • 7 / 2

Start / End Page

  • 108 - 111

PubMed ID

  • 7382914

International Standard Serial Number (ISSN)

  • 0094-2405

Digital Object Identifier (DOI)

  • 10.1118/1.594673

Language

  • eng

Conference Location