6D image guidance for spinal non-invasive stereotactic body radiation therapy: Comparison between ExacTrac X-ray 6D with kilo-voltage cone-beam CT.

Published

Journal Article

PURPOSE: To investigate setup discrepancies measured with ExacTrac X-ray 6 degree-of-freedom (6D) and cone-beam computed tomography (CBCT) for patients under treatments of stereotactic body radiation therapy (SBRT). MATERIALS AND METHODS: In this work, phantom and patient studies were performed. In the phantom studies, an anthropomorphic phantom was placed with pre-defined positions, and imaged with ExacTrac X-ray 6D and CBCT to test the accuracy of the imaging systems. In the patient studies, 16 spinal SBRT patient cases were retrospectively analyzed. The patients were initially positioned in customized immobilization cradles and then aligned with ExacTrac X-ray 6D and CBCT. The setup discrepancies were computed and quantitatively analyzed. RESULTS: This study indicates modest discrepancies between ExacTrac X-ray 6D and CBCT with spinal SBRT. The phantom experiments showed that translational and rotational discrepancies in root-mean-square (RMS) between ExacTrac X-ray 6D and CBCT were, respectively, <1.0mm and <1 degrees . In the retrospective patient studies, translational and rotational discrepancies in RMS between ExacTrac X-ray 6D and CBCT were <2.0mm and <1.5 degrees . CONCLUSIONS: ExacTrac X-ray 6D represents a potential alternative to CBCT; however, pre-caution should be taken when only ExacTrac X-ray 6D is used to guide SBRT with small setup margins.

Full Text

Duke Authors

Cited Authors

  • Chang, Z; Wang, Z; Ma, J; O'Daniel, JC; Kirkpatrick, J; Yin, F-F

Published Date

  • April 2010

Published In

Volume / Issue

  • 95 / 1

Start / End Page

  • 116 - 121

PubMed ID

  • 20122747

Pubmed Central ID

  • 20122747

Electronic International Standard Serial Number (EISSN)

  • 1879-0887

Digital Object Identifier (DOI)

  • 10.1016/j.radonc.2009.12.036

Language

  • eng

Conference Location

  • Ireland