Design considerations and initial performance of a 1.2cm2 beta imaging intra-operative probe

Published

Journal Article

A novel small area beta detector is under development for nuclear emission imaging of surgically exposed radiolabeled tumor beds. The detector size (1.2 cm2) is determined by the requirement to introduce the probe into small cavities, e.g. during neuro-surgical lesion resection, but large enough to produce images of clinical significance. The various components of the imaging chain are optimized by both simulation and measurements.

Duke Authors

Cited Authors

  • Tornai, MP; MacDonald, LR; Levin, CS; Siegel, S; Hoffman, EJ

Published Date

  • December 1, 1995

Published In

  • Ieee Nuclear Science Symposium &Amp; Medical Imaging Conference

Volume / Issue

  • 3 /

Start / End Page

  • 1791 - 1795

Citation Source

  • Scopus