Investigation of the resolution characteristics of a rotating parallel hole collimator for use in a medium energy SPECT system

Published

Journal Article

A line source filled with 0.4-1.3 mCi of I-31 was imaged on a SPECT system. The rotating hole collimator and a medium energy collimator were placed on adjacent heads of the camera. Results indicate that with proper rotation, there will be no degradation of the SPECT image from the rotating collimator. For the case of the 9 cm radius of rotation, the rotating parallel hole collimator's full width at half maximum was measured to be 5.2-5.8 mm.

Duke Authors

Cited Authors

  • Stone, CD; Jaszczak, RJ; Smith, MF; Greer, KL; Bowsher, JE

Published Date

  • January 1, 1999

Published In

  • Ieee Nuclear Science Symposium and Medical Imaging Conference

Volume / Issue

  • 2 /

Start / End Page

  • 1305 - 1311

Citation Source

  • Scopus