Low-cost, high-performance array detector for spectroscopy based on a charge-coupled photodiode

Published

Journal Article

We describe a nitrogen-cooled array detector system based on the Thomson TH 7832A charge-coupled photodiode. Performance comparisons are made between this system and others based on two-dimensional CCDs. The TH 7832A compares favorably in resolution, spectral range, and readout speed. Analysis of read and shot noise contributions identifies classes of experiments for which the larger read noise of the TH 7832A is inconsequential.

Full Text

Duke Authors

Cited Authors

  • Johnson, JB; Edwards, G; Mendenhall, M

Published Date

  • December 1, 1994

Published In

Volume / Issue

  • 65 / 5

Start / End Page

  • 1782 - 1783

International Standard Serial Number (ISSN)

  • 0034-6748

Digital Object Identifier (DOI)

  • 10.1063/1.1144826

Citation Source

  • Scopus