Needle-based reflection refractometry of scattering samples using coherence-gated detection

Published

Journal Article

We present a novel method for in situ refractive index measurement of scattering samples using a needle device. The device employs a fiber-based reflectance refractometer and coherence-gated detection of the reflected optical signal that eliminates scattering-dependent backreflection contributions. Additionally, birefringence changes induced by fiber movement are neutralized by randomizing the source polarizations and averaging the measured Fresnel reflection coefficients over many incident polarization states. Experimental measurements of Intralipid scattering solutions are presented and compared with Monte Carlo simulations. © 2007 Optical Society of America.

Full Text

Cited Authors

  • Zysk, AM; Marks, DL; Liu, DY; Boppart, SA

Published Date

  • April 16, 2007

Published In

Volume / Issue

  • 15 / 8

Start / End Page

  • 4787 - 4794

Electronic International Standard Serial Number (EISSN)

  • 1094-4087

Digital Object Identifier (DOI)

  • 10.1364/OE.15.004787

Citation Source

  • Scopus