Investigating the effects of energy resolution in dedicated emission mammotomography

Publication ,  Journal Article
Cutler, SJ; Brzymialkiewicz, CN; Tornai, MP
Published in: IEEE Nuclear Science Symposium Conference Record
December 1, 2005

This study probes the recent debate over the necessity for good energy resolution for uncompressed breast, 3D lesion imaging with dedicated single photon emission mammotomography. Here, the imaging system consists of a commercial, discretized CZT gamma camera having ∼6% FWHM intrinsic energy resolution (at 140keV) and intrinsic spatial resolution corresponding to the 2.5mm square pixilation, and is used on a fully-3D positioning gantry. Wider energy windows are used on list mode acquired data as a surrogate for having otherwise identical detection systems with poorer energy resolution characteristics. Scans using simple circular trajectories are first obtained of an aqueous Tc-99m filled mini resolution cold-rod phantom at various radii-of-rotation, and also immersed in a larger uniform water bath. Multiple 3D orbits about Tc-99m filled anthropomorphic breast and torso phantoms are acquired, with the breast containing two large lesions. The list mode data files were multiply processed to obtain images of varying energy window widths (from symmetric 6% to an asymmetric 18% (-12+6)) but with the same projection image count density. Counts were randomly subsampled from the entire list mode data set in order to maintain equivalent levels of count density for several bootstrap realizations. All data was then reconstructed using OSEM for various iterations. Profiles were obtained from the cold rod images, and regions of interest were drawn in and about the spherical lesions to determine signal-tonoise ratios and contrasts for each iteration. Results clearly illustrate both visual and quantitative differences between the various energy windows, with smaller energy windows (corresponding to better energy resolution) having better image quality. © 2005 IEEE.

Duke Scholars

Published In

IEEE Nuclear Science Symposium Conference Record

DOI

ISSN

1095-7863

Publication Date

December 1, 2005

Volume

5

Start / End Page

2537 / 2541
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Cutler, S. J., Brzymialkiewicz, C. N., & Tornai, M. P. (2005). Investigating the effects of energy resolution in dedicated emission mammotomography. IEEE Nuclear Science Symposium Conference Record, 5, 2537–2541. https://doi.org/10.1109/NSSMIC.2005.1596856
Cutler, S. J., C. N. Brzymialkiewicz, and M. P. Tornai. “Investigating the effects of energy resolution in dedicated emission mammotomography.” IEEE Nuclear Science Symposium Conference Record 5 (December 1, 2005): 2537–41. https://doi.org/10.1109/NSSMIC.2005.1596856.
Cutler SJ, Brzymialkiewicz CN, Tornai MP. Investigating the effects of energy resolution in dedicated emission mammotomography. IEEE Nuclear Science Symposium Conference Record. 2005 Dec 1;5:2537–41.
Cutler, S. J., et al. “Investigating the effects of energy resolution in dedicated emission mammotomography.” IEEE Nuclear Science Symposium Conference Record, vol. 5, Dec. 2005, pp. 2537–41. Scopus, doi:10.1109/NSSMIC.2005.1596856.
Cutler SJ, Brzymialkiewicz CN, Tornai MP. Investigating the effects of energy resolution in dedicated emission mammotomography. IEEE Nuclear Science Symposium Conference Record. 2005 Dec 1;5:2537–2541.

Published In

IEEE Nuclear Science Symposium Conference Record

DOI

ISSN

1095-7863

Publication Date

December 1, 2005

Volume

5

Start / End Page

2537 / 2541