4D CT image sets are used to encode patient-organ motion and to calculate the dose delivered during each respiratory phase. The dose-per-phase is summarised to the total dose via deformable registration. The system incorporates the actual beam parameters and synchronises the beam motion (e.g., gantry/MLC motion of the DARC and MLC motion of the IMRT) to the patient's respiratory motion. Furthermore, this system incorporates different treatment techniques such as 3D conformal (3DCRT), dynamic arc (DARC), and Intensity-Modulated Radiation Therapy (IMRT), thus allows better understanding of organ motion effect on dose delivery, treatment techniques and corresponding optimal margins. Copyright © 2012 Inderscience Enterprises Ltd.