Combined doppler and phased-array echocardiographic estimation of cardiac output.

Journal Article (Journal Article)

The capability of a pulsed Doppler flowmeter combined with a phased-array imaging system to measure volume flow was tested in vitro and in patients undergoing cardiac catheterization. The Doppler-phased-array system (DPA) was used to determine vessel diameter and a superimposed cursor was used to locate the range and angle of the Doppler sample volume. DPA estimates of continuous flow through tubing in a water tank correlated strongly (r = 0.99) with measured flow corresponding to physiologic ranges from 3-12 l/min. For pulsatile flow in a water tank, a correlation of r = 0.86 with measured flow was obtained, whereas DPA estimates of cardiac outputs compared with Fick estimates in the 11 patients produced a correlation of r = 0.83. These data indicate that estimates of cardiac output are possible using the DPA approach.

Full Text

Duke Authors

Cited Authors

  • Magnin, PA; Stewart, JA; Myers, S; von Ramm, O; Kisslo, JA

Published Date

  • February 1, 1981

Published In

Volume / Issue

  • 63 / 2

Start / End Page

  • 388 - 392

PubMed ID

  • 7449060

International Standard Serial Number (ISSN)

  • 0009-7322

Digital Object Identifier (DOI)

  • 10.1161/01.cir.63.2.388

Language

  • eng

Conference Location

  • United States