Skip to main content
Journal cover image

Polyethylene liner exchange of the Harris-Galante porous I and II acetabular components without cement: results and complications.

Publication ,  Journal Article
Lachiewicz, PF; Soileau, ES
Published in: J Arthroplasty
October 2006

This study reviewed the long-term efficacy of the locking mechanism of the Harris-Galante porous (HGP) I and II acetabular components and the results and complications of polyethylene liner exchange without cement. There were 400 HGP-I components with a mean follow-up of 10 years (range, 2-19 years) and 78 HGP-II components with 8 years of mean follow-up (range, 2-13 years). There has been only 1 liner dislodgement (0.2%). Thirty-five hips (34 patients) have undergone liner exchange without cement. The index acetabular component was implanted as a primary procedure in 19 hips and a revision in 16 hips. No exchanged liner has dislodged at a mean follow-up time of 5.1 years (range, 2-11 years). However, there have been 7 patients (20%) with recurrent dislocation and all required reoperation. Dislocation was significantly lower when an elevated rim liner was used.

Duke Scholars

Published In

J Arthroplasty

DOI

ISSN

0883-5403

Publication Date

October 2006

Volume

21

Issue

7

Start / End Page

992 / 997

Location

United States

Related Subject Headings

  • Reoperation
  • Prosthesis Failure
  • Prosthesis Design
  • Porosity
  • Polyethylene
  • Orthopedics
  • Middle Aged
  • Male
  • Humans
  • Follow-Up Studies
 

Citation

APA
Chicago
ICMJE
MLA
NLM
Lachiewicz, P. F., & Soileau, E. S. (2006). Polyethylene liner exchange of the Harris-Galante porous I and II acetabular components without cement: results and complications. J Arthroplasty, 21(7), 992–997. https://doi.org/10.1016/j.arth.2005.06.019
Lachiewicz, Paul F., and Elizabeth S. Soileau. “Polyethylene liner exchange of the Harris-Galante porous I and II acetabular components without cement: results and complications.J Arthroplasty 21, no. 7 (October 2006): 992–97. https://doi.org/10.1016/j.arth.2005.06.019.
Lachiewicz, Paul F., and Elizabeth S. Soileau. “Polyethylene liner exchange of the Harris-Galante porous I and II acetabular components without cement: results and complications.J Arthroplasty, vol. 21, no. 7, Oct. 2006, pp. 992–97. Pubmed, doi:10.1016/j.arth.2005.06.019.
Journal cover image

Published In

J Arthroplasty

DOI

ISSN

0883-5403

Publication Date

October 2006

Volume

21

Issue

7

Start / End Page

992 / 997

Location

United States

Related Subject Headings

  • Reoperation
  • Prosthesis Failure
  • Prosthesis Design
  • Porosity
  • Polyethylene
  • Orthopedics
  • Middle Aged
  • Male
  • Humans
  • Follow-Up Studies