16-channel neural pre-conditioning device

Published

Conference Paper

© 2003 IEEE. We present the mixed-signal circuit design, layout, implementation techniques, and test data for a 16-channel neural pre-conditioning device that is used to amplify and filter signals acquired from chronically implanted electrodes in an animal's brain. Schematics and simulation data for each of the subcircuit macros are presented which include a high gain, continuous time first order bandpass filter pre-amplifier, a cascaded bandpass switch capacitor filter, a selectable gain output buffer, and a voltage controlled oscillator based clock generation circuitry. This device was implemented using AMI's, 0.5 μm, double poly, triple level metal, 5 V, CMOS technology. The layout and floorplan, specifications and test data for this device conclude this paper.

Full Text

Duke Authors

Cited Authors

  • Morizio, J; Won, D; Obeid, I; Bossetti, C; Nicolelis, M; Wolf, P

Published Date

  • January 1, 2003

Published In

Volume / Issue

  • 2003-January /

Start / End Page

  • 104 - 107

Electronic International Standard Serial Number (EISSN)

  • 1948-3554

International Standard Serial Number (ISSN)

  • 1948-3546

International Standard Book Number 10 (ISBN-10)

  • 0780375793

Digital Object Identifier (DOI)

  • 10.1109/CNE.2003.1196767

Citation Source

  • Scopus